DGS Regular maintenance day(5/30)
Confirmation during O4a
Stepper motor
[Checked]
Check if the power supply can be controlled by BIO
Check if the script can be started
Checked that the limit switches installed are green.
[Check result]
Lock loss when turned on, also no off-loading this week, so no confirmation is foregone.
Release Sensor
[Checked the Release Sensor.]
Checked that Release Sensor is 0, 1: Seated.
[Check result]
ETMX, ETMY, ITMX, ITMY : No problem
Picomotor
[Confirmation]
PingTest is performed.
[Result]
Only known
NG:
MCI,STM1,
POS_POM,POM1
=> Confirmation only, no recovery operation was performed.
OK:
ETMX,ETMY,ITMX,ITMY
BS(IM,BF),SRM(IM,BF),SR2(IM,BF),SR3(IM,BF,STM)
PRM(IM,BF),PR2(IM,BF),PR3(IM,BF)
MCE,IMMT1,IMMT2,OMMT1,OMMT2,OSTM
REFL_WFS,POP,POP2,POS,POS2,POP_WFS,AS_WFS
POP_POM, MCF(MCO,MCI)
PCAL(EX1,EX2,EY1,EY2):Not checked because it is ON only when used
Temperature controller
[Checked]
Is the temperature acquisition working properly?
[Check result]
Is the temperature acquisition working properly?
=> No problem
Precision air Processer
[Check Contents]
Is the temperature acquisition working properly?
[Result]
No problem
FunctionGenerator
(Confirmation)
Is there a ping response?
Are all EPICS channels alive?
[sitemap]-[Commissioning]-[ALS FG]-[Xarm],[Yarm].
[Check result]
X and Y PLLs are working properly.
HWP control PC
(Confirmation)
Check if it is possible to login to the control PC.
(Result)
Both k1hwp0, 1 , 2, 3 and 4 are OK.