Reports of 32931
MIF (General)
takahiro.yamamoto - 21:05 Friday 07 November 2025 (35531) Print this report
The Return of the Combs

For checking that the frequency points at the swept sine of calibration measurements are contaminated by spectral peaks, I made a latest ASD with fine (~mHz) resolution.
Then I found that 4.17Hz and 9.77Hz combs came back which were seen also on OMC dark and CARM error signal, respectively as shown in Fig.1 and Fig.2.

These two plots were made with the data around 15:00 JST and 23:00 JST on Nov. 6th. Former and latter locks can be seen in Fig.3 as t = 6h-11h and t = 14h-20.5h, respectively. The lock duration of these two were ~5hr and ~6.5hr and there was no human activities in order to check the stability with disabling the whitening filter stage for the DCPD (probably for former lock and surely for latter lock in my memory).

These two combs were loud during O4a and were mitigated before before O4c (Comparison around 100Hz would be the easiest to understand). I couldn't remember how to mitigate the 4.17Hz combs on OMC (pin assignment of ZSW, cabling route around AS table, or else?). Though there was possibility that noise itself became louder than before, disabling the whitening filter makes SNR of these series peaks louder. If so, there may be no way to suppress them for now because stable IFO is unavailable with engaging whitening filter. In my memory about 9.77Hz combs on CARM, they were able to mitigated by adjusting the input gain of CMS as positive. Though I was not able to track recent activities about replacing IMC-CMS precisely, now the original common mode servo chassis were used for IMC and input gain were set as positive. So I have no idea why 9.77Hz combs came back. replacing chassis and/or re-cabling changed some situation around electrical connections...?

Images attached to this report
MIF (General)
takafumi.ushiba - 18:02 Friday 07 November 2025 (35530) Print this report
Comment to Frequent kick of suspensions (35505)

I checked the OMC DC PD signals in each lock from last night.

Figure 1 shows the OMC DC PD signals at the lock from 2025/11/6 06:03:48 UTC.
Since the speed of glitches is very fast, the signals would be enhanced by a factor of 10 if the whitening filter is off.
So, if the glitch reached around upper Y-cursor, the signals would be saturated if the whitening filter is on.
There are four glitches that exceeds the upper Y cursor, so the lockloss should be happened if the whitening filter was turned on.

Figure 2 shows the enlarged view of one glitch in fig1.
Clear ring down can be seen, so it seems due to the kick of suspensions.

For the lock from 2025/11/6 11:41:06 UTC, these is no large glitch that exceeds the threshold.
For the lock from 2025/11/6 12:22:33 UTC, these is also no large glitch that exceeds the threshold but the glitches close to te threshld can be seen as shown in fig3.

For the lock from 2025/11/6 13:55:08 UTC, there is 2 glitches that exceed the thresold.
One of them is smilar glitch as shown in fig2 but the other is different: very short glitch calmed down within 1 second (fig4).
The similar short glitch can be seen in the lock from 2025/11/6 21:48:58 UTC, which seems very clear sine gaussian noise (fig5).

So, some DC PD saturation seemed suspension kick but there might be another reason of sudden glitch in OMC DC PD.

Images attached to this comment
OBS (Summary)
tatsuki.washimi - 17:24 Friday 07 November 2025 (35529) Print this report
Operation Shift Summary (Nov. 7, 2025)

Operator: YokozaWashimi
Shift time: 9:00-17:00 JST

Check Items:
VAC: No issues were found.
CRY temp: No issues were found.
CRY cooling water: No issues were found.
Temp: No issues were found.
VIS GAS filter: No issues were found.

IFO state(JST):
9:00 : STANDBY
17:00 : CALIB NOT READY

FCL (Electricity)
masakazu.aoumi - 15:59 Friday 07 November 2025 (35528) Print this report
Monthly inspection of electric equipments
Monthly inspection of electric equipments
With Nakai-denki san
13:00 In
15:15 Out
DetChar (General)
hirotaka.yuzurihara - 15:37 Friday 07 November 2025 (35527) Print this report
Comment to Temporary suspension of uploading segment files from Kamioka to DQSEGDB (34837)

This is a report of last week. Sorry for late.

As reported in klog34823, the gap of the frame data in Kashiwa cluster (called system-B) was filled on 10/10. So, I regenerated the segment files of 8/13 data, via k1det2.
The working directory is /users/yuzu/work/20251030_segment@k1det1. I used the conda environment of igwn-py38, which is same environment as the regular segment production.
I asked the final review of the generated segment files to YamaT-san. After his confirmation, I will copy the files to the /users/DET/Segments and ask the upload of segment files to DQSEGDB.

NOTE

  • To reproduce the segment files reading 1 day data, it took around 30 minutes.
  • After the trouble of the missing frame data, I just suspended the copy process from the temporal directory to /users/DET/Segments. So, the other segment files (after 8/13) are regularly generated and stored in /users/DET/tools/Segments/Script/Partial.
  • From August 15 (in UTC timezone), the interferometer was unfortunately not in the OBSERVATION state due to the laser issue (klog34839 and klog34840). So, we don't need to care the segment files after August 15.
MIF (General)
takafumi.ushiba - 14:01 Friday 07 November 2025 (35525) Print this report
MICH and PRCL FF optimization

Abstract:

It doesn't seem easy to improve the MICH and PRCL FF performance by changing the gain.
So, I keep current setting for them.

Detail:

I measured the Tf from MICH OUT to DARM to optimize MICH FF gain.
Figure 1 shows the measured TF with various condition.

Blue line shows the reference, which was optimized just before joining O4c.
Green line shows the current TF before optimization.

To improve the MICH FF performance, a gain of 1.11 is better than the gain of 1 (compared with brown and pink line in fig1).
However, even though changing the MICH FF gain to 1.11, overall performance (after engaging PRCL FF at the same time) of MICH FF becomes worse (compared with green and cyan lines).
I also tried to optimize MICH2PRCL FF and PRCL2DARM FF but it is hard to find the good configuration with a short term.

So, I keep FF setting as it was.

Images attached to this report
DetChar (General)
takahiro.yamamoto - 13:56 Friday 07 November 2025 (35526) Print this report
Maintenance of SummaryPages server
It's a part of preparation for migrating from the LVK era to IGWN era.

Settings of network, firewall, and some daemon processes on the SummaryPages server were updated.
New DQSEGDB servers with IGWN domain are added to an allow list of firewalld and rsyncd daemon.
Firewall of ICRR DMZ had been already updated by the administrator.
MIF (General)
takaaki.yokozawa - 12:03 Friday 07 November 2025 (35521) Print this report
Lock loss investigation 251106 night - 251107 morning
[Yuzu, Yokozawa]

We performed the investigation of the lock loss last night.

Compared from yesterday, we cannot see the lock loss due to the ETMX kick.

Lock loss lists
# 1562 2025-11-06 10:54:57.562500 UTC ( 17468s)
# 1563 2025-11-06 11:48:24.687500 UTC ( 438s)
# 1564 2025-11-06 13:23:24.937500 UTC ( 3650s)
# 1565 2025-11-06 20:30:29.812500 UTC ( 23721s)
# 1566 2025-11-06 21:20:26.937500 UTC ( 1408s)

ETMY MN oplev glitch
#1565, #1566

Nearby Earthquake
#1564

PRCL control signal became strange.
#1562, #1563
We checked more carefully about those locked losss
Fig.1. showed the PRCL LSC related channels, obviously, the signal saturation detected in POP RF45 I,Q (RFPD signal for the PRCL 1f lock)
We don't know the origin of this glitch, it happened several times (Other time in Fig.2. and Fig.3.)
Images attached to this report
FCL (Water)
takashi.uchiyama - 12:02 Friday 07 November 2025 (35524) Print this report
Comment to Water pump alert at Yend (35331)
2025/11/07

Tonetsu, Hayakawa, Kimura, Yasui, Uchiyama

We performed the repair work. We can operate both No. 1 anf No. 2 pump alternatvely.
In this repair work, we replaced the float switch of both No. 1 and No. 2 pump.
OBS (SDF)
dan.chen - 11:42 Friday 07 November 2025 (35523) Print this report
Comment to Changes of observation.snap during O4c (34169)

I accepted Pcal beam position related SDF diffs as reported klog35522.

k1calex

k1caley

Images attached to this comment
CAL (Pcal general)
dan.chen - 11:38 Friday 07 November 2025 (35522) Print this report
Pcal Parameter Update Report

A CAL Tcam session was performed to obtain beam position information necessary for Pcal. The parameters have already been updated, and SDF has been accepted.

Operator: Dan Chen

Update Time: 2025/11/07 11:27:28

EPICS Key Before [mm] After [mm] Δ (After - Before) [mm]
K1:CAL-PCAL_EX_TCAM_PATH1_X 0.33702 mm 0.33957 mm +0.00255 mm
K1:CAL-PCAL_EX_TCAM_PATH1_Y 67.43505 mm 67.14666 mm -0.28839 mm
K1:CAL-PCAL_EX_TCAM_PATH2_X -0.10011 mm -0.19262 mm -0.09251 mm
K1:CAL-PCAL_EX_TCAM_PATH2_Y -63.12901 mm -63.44511 mm -0.31610 mm

Update Time: 2025/11/07 11:28:31

EPICS Key Before [mm] After [mm] Δ (After - Before) [mm]
K1:CAL-PCAL_EX_TCAM_MAIN_X 4.14543 mm 4.10822 mm -0.03721 mm
K1:CAL-PCAL_EX_TCAM_MAIN_Y 12.29349 mm 13.23050 mm +0.93701 mm

Update Time: 2025/11/07 11:29:07

EPICS Key Before [mm] After [mm] Δ (After - Before) [mm]
K1:CAL-PCAL_EY_TCAM_PATH1_X 1.07761 mm 1.08398 mm +0.00637 mm
K1:CAL-PCAL_EY_TCAM_PATH1_Y 64.59614 mm 65.04131 mm +0.44517 mm
K1:CAL-PCAL_EY_TCAM_PATH2_X -0.72208 mm -0.93095 mm -0.20887 mm
K1:CAL-PCAL_EY_TCAM_PATH2_Y -70.16760 mm -69.74364 mm +0.42396 mm

Update Time: 2025/11/07 11:29:38

EPICS Key Before [mm] After [mm] Δ (After - Before) [mm]
K1:CAL-PCAL_EY_TCAM_MAIN_X 9.47443 mm 5.33197 mm -4.14245 mm
K1:CAL-PCAL_EY_TCAM_MAIN_Y -3.65031 mm -2.29115 mm +1.35917 mm

 

OBS (SDF)
takafumi.ushiba - 11:21 Friday 07 November 2025 (35519) Print this report
Comment to Changes of observation.snap during O4c (34169)

I accepted the following SDFs.

k1asc1 (fig1):

These changes were done for IFO stability check on November 6.
Since TRAMP doesn't affect to the IFO condition, I accepted them.

k1lsc (fig2):

All changes are related to the work reported in klog35505.
Since they should not affect the IFO condition, they are accepted.

k1omc (fig3):

All changes are related to the work reported in klog35510.
Since these changes are for improving the stability of IFO,  they should be accepted.

k1visetmxp (fig4):

All changes are related to the work reported in klog35510.
Since they should not affect the IFO condition, they are accepted.

k1ascbpc (fig5):

This change is related to the work reported in klog35504.
Since this change is for keeping beam spot on ETMY at the center, it should be accepted.

Images attached to this comment
LAS (General)
shinji.miyoki - 10:55 Friday 07 November 2025 (35520) Print this report
Comment to Innolight Mephisto2000NE Inside (35461)

The followings are the memo related with cleaning (2009Feb02).

--- Cleaning ---
We performed cleaning inside the Mephisto2000NE.
Result: Power increased from 800mW (2.3A) to 2W (2.3A).

--- Why does the interior get so dirty? ---
First, the booth cleanliness is ound Class 300, which should be acceptable. The HEPA filter operates at AC60V.

According to Mr. Nishibata (Indeco Inc at that time):
1) Lasers used by customers for applications heavily utilizing PZT tend to get dirty quickly elsewhere too. On the other hand, those not using PZT show slower fouling.
2) The crystal surface facing the PZT wiring is overwhelmingly dirty. In other words, it's likely that vibrating the PZT causes dust to be ejected from the wiring and adhere to the crystal surface. Since this is an internal structural issue, I requested that it be addressed in the new laser currently under production. For the current laser, covering the PZT wiring with Kapton tape might help, but we haven't done that yet. 

CRY (General)
shoichi.oshino - 10:02 Friday 07 November 2025 (35518) Print this report
Modified CRY_MON guardian
As requested by Kimura-san, I modified the CRY_MON guardian to send messages according to the respective channel when an alert occurs.
In the CRY_MON guardian code, I changed the list of pump related channels to a dictionary and added a clear warning message for each channel.
I confirmed that these changes have been reflected and that the intended warning message is sent to Slack.
CAL (General)
hirotaka.yuzurihara - 8:54 Friday 07 November 2025 (35517) Print this report
TCam photo session 20251107

I took the TCam photos for commissioning at 08:43 ~ 08:50 this morning. The previous work is klog35448. We don't need to update the reference positions.

CRY (General)
shinji.miyoki - 8:34 Friday 07 November 2025 (35516) Print this report
Comment to IY, EX Im heater Adjustment (35502)

So, so stable IM temperatures for IY and EX, while EY is increasing.

Images attached to this comment
FCL (Air)
shinji.miyoki - 7:54 Friday 07 November 2025 (35515) Print this report
temp adjustment in the corner station

Since the temperatures in the corner station appear to be decreasing, I increased the target temperature of the SR-C precision cooler from 18.00 to 20.00 around 6:30.

Images attached to this report
CRY (General)
shinji.miyoki - 6:10 Friday 07 November 2025 (35514) Print this report
Comment to IY, EX Im heater Adjustment (35502)

I slightly adjusted currents for EX/IY (Fig.1, 2) IMs around 0:00 and 6:00 today. If the suspension kicking rate will not change, we should consider other reasons, except for the temperature changes in the cryostat area.

Images attached to this comment
CRY (General)
shinji.miyoki - 23:15 Thursday 06 November 2025 (35513) Print this report
Comment to IY, EX Im heater Adjustment (35502)

I increased the current for IY IM to realize the constant condition around ~40K. The temp of EX IM seemed to be around 40K.

Images attached to this comment
CRY (General)
shinji.miyoki - 18:49 Thursday 06 November 2025 (35512) Print this report
Comment to IY, EX Im heater Adjustment (35502)

I increased the current for EX IM to realize the constant condition around ~40K.

Images attached to this comment
FCL (Air)
shinji.miyoki - 18:28 Thursday 06 November 2025 (35511) Print this report
IXV temp adjustment

Because the temperature in IXV decreased and F0 GAS position is also decreasing, i increased the Delongh heater power from 600W to 900W around 18:30.

F2 is also decreasing.

Images attached to this report
MIF (General)
takafumi.ushiba - 17:20 Thursday 06 November 2025 (35510) Print this report
Comment to Frequent kick of suspensions (35505)

To check if the reason of OMC DC PD saturation is the suspension kick, I locked PRFPMI without engaging whitening filter for OMC DC PDs.
It degrades the ADC noise and noise floor got slightly worse, ADC range at high frequency increases by a factor of 10.
So, if the large suspension kick is the reason of the lockloss, it would help t keep the lock.
To keep this configuration, I modified the OMC_LSC guardian as shown in fig1.

Images attached to this comment
OBS (Summary)
tatsuki.washimi - 17:12 Thursday 06 November 2025 (35509) Print this report
Operation Shift Summary (Nov. 6, 2025)

Operator: YokozaWashimi
Shift time: 9:00-17:00 JST

Check Items:
VAC: No issues were found.
CRY temp: No issues were found.
CRY cooling water: No issues were found.
Temp: No issues were found.
VIS GAS filter: No issues were found.

IFO state(JST):
9:00 : LOCKING
17:00 : LOCKED

AEL (General)
masakazu.aoumi - 15:53 Thursday 06 November 2025 (35508) Print this report
New IRIG-B Functionality Verification Test
New IRIG-B Functionality Verification Test
SK Test Bench

Reconnect the IRIG-B signal from the computer currently connected to IRIG-B (k1ex1-dev k1iopmcf0) to the new IRIG-B (S2516844). Verify functionality after restarting.

Verify synchronization in K1IOPMCF0_GDS_TP.adl -> Synchronization is achieved
Compare k1dc0's GPS[s] and k1iopmcf0's GPS[s] in K1CDS-DAQ_OVERVIEW.adl -> Count values match

Future Schedule
Measure timing of IRIG signals between current IRIG-B and new IRIG-B


新IRIG-B動作確認テスト
SK テストベンチ

現行動作中のIRIG-Bに接続されているコンピュータ(k1ex1-dev k1iopmcf0)のIRIG-B信号を新IRIG-B(S2516844)に接続し直し、再起動後動作確認をする。

K1IOPMCF0_GDS_TP.adlで同期がとれていることを確認->同期は取れている
K1CDS-DAQ_OVERVIEW.adlでk1dc0のGPS[s]とk1iopmcf0のGPS[s]を比較->カウント数は同じ

今後の予定
現行IRIG-Bと新IRIG-BのIRIG信号のタイミング計測
CRY (General)
shinji.miyoki - 14:54 Thursday 06 November 2025 (35507) Print this report
Comment to IY, EX Im heater Adjustment (35502)

I increased the current for EX IM.

Images attached to this comment
Search Help
×

Warning

×