takahiro.yamamoto - 18:01 Monday 20 October 2025 (35380) Print this report
Memo for recovering calibration
Changes since 18:00 on Aug. 8th (last weekly calibration) to 11:30 on Oct. 20th.

foton filters
  • k1lsc
  • k1asc1
    • klog#35180: DHARD_P_FM7, DHARD_Y_FM3, CHARD_P_FM7, CHARD_Y_FM8, DHARD_P_FM1, DHARD_Y_FM9, CHARD_P_FM1, CHARD_Y_FM6, CSOFT_Y_FM8
    • klog#35187: MICH_P_FM4, MICH_Y_{FM2,FM3}, PRC2_P_{FM2,FM3}, PRC1_Y_FM5
    • klog#35193: PRC2_Y_FM2
    • klog#35222: PRC2_Y_FM3
    • Cannot find: PRC1_Y_FM7, CSOFT_P_FM6
  • k1ial
    • Cannot find: X_PNC_SERVO_{FM3,FM4}
  • k1imc
    • klog#34980: SERVO_NPRO_TEMP_BIAS_FM2 (Not loaded yet)
  • k1visetmxlsc


SDF tables (Only contains accepted diffs and lots of unaccepted diffs still remain)

guardian codes
  • IMC.py
  • PLL.py
  • LSC_LOCK.py
    • klog#35093: Adjust IN2 gain of CARM-CMS
    • klog#35103: Adjust HWP angle, Change REFL threshold for power up and Increase DARM offset
    • klog#35127: Tweak IN2 gain of CARM-CMS
    • klog#35144: Tweak IN2 gain of CARM-CMS
    • klog#35227: Add tidal offload
    • klog#35357: Adjust HWP angle
    • Cannot find: Reduce POP90 threshold
    • Cannot find: Increase DARM gain for DC readout at INMTRX (*2)
  • PMC.py
  • ASC_LOCK.py
    • Cannot find: Just added test code and commented out now(?)
  • VERTEX.py
    • Cannot find: Reduce POP90 threshold


*1) Because gain reduction from the actuator to the sensor was compensated by the servo gains, a sum of sensing and actuation paths is now incorrect for MICH and PRCL.
*2) If INMTRX gain compensates reduction of the optical gain, DARM calibration is kept within accuracy of gain estimation. If not, a sum of sensing and actuation paths is now incorrect for DARM.
Comments to this report:
takahiro.yamamoto - 18:03 Tuesday 04 November 2025 (35479) Print this report
Changes since 11:30 on Oct. 20th (last weekly calibration) to 15:00 on Nov. 4th.

foton filters

SDF tables (Lots of unaccepted differences still remains)

guardian codes
  • CRY_MON.py
    • Cannot find: Update channel list
  • IMC.py
    • klog#35394: Change IN1/IN2 gain, SLOWFILTER, and FASTPOL for modified CMS (*5)
    • Cannot find: Change setpoint of PSL_HWP
    • klog#35449: Revert IN1/IN2 gain for original CMS
  • PLL.py
    • Cannot find: Set different minimum temp bias for X/Y
  • PMC.py
    • Cannot find: Adjust common gain for NeoLASE
  • LSC_LOCK.py
    • Cannot find: Change threshold for ADS
    • klog#35475: Change setpoint for REFL_HWP
    • klog#35475: Change threshold for OMC IAL


*1) They doesn't affect calibration TF models because they are probably used only for some measurement and turned off as default.
*2) If it's a compensation factor of a change in the optical gain, overall gain of current DARM calibration should not so bad estimation. It must be updated as simultaneous updates of the optical gain (CAL-CS_DARM_ERR_FM1) and inverse sensing function (CAL-CS_DARM_ERR_FM9)
*3) If it's a compensation factor of a change in the optical gain, overall gain of current MICH/PRCL calibration are wrong as 2-3dB.
*4) It affects actuation function of DARM and notch (rapid phase change) near the UGF probably makes large calibration error like O4a.
*5) Though SLOWFILTER and FASTPOL were changed according to the guardian code, SLOWFILTER and SLOWPOL should be changed for keeping overall polarity...? Is it my misunderstanding?