Reports 1-1 of 1 Clear search Modify search
IOO (General)
takafumi.ushiba - 12:58 Thursday 16 October 2025 (35352) Print this report
Comment to PMC, IMC, and PLLs were locked with the 3rd fiber amp. laser (35340)

Since nominal IMC CMS IN1 gain was changed due to the work reported in original post, I changed IMC CMS IN2 gain and LSC-MCL servo.
I changed IMC CMS IN2 gain from 7dB to 2dB because IN1 gain was changed frm 23dB to 18dB.
For MCL, I reduced the gain by -1.8dB so that crossover frequency of MCL keeps 100Hz.
Figure 1 shows the crossover frequency measurement result after gain adjustment, which seems fine.