Replacement IMC servo and EOM amplifier

Ushiba, Tanaka

Abstract

We repleced IMC CMS to gain modified one and also replaced EOM amlifier to x20 gain one in order to enlarge the actuator range. For now, IMC seems to keep the lock for 2 hours. We left IMC lock to check the stability. 

What we did

Replacement IMC CMS to the modified one

We brought a modified IMC CMS by Moriwaki-san to a mine. Moriwaki-san modified a gain of a slow gain filter in a SLOW PATH of IMC CMS from -0.1 to 1 and also modified a gain in FAST PATH of IMC CMS from 1 to 10 (The detail modification is reported in klog35396).

Before the replacement, we confirmed some TFs of the original IMC CMS. Fig. 1 and Fig. 2 show OLTF of the IMC loop and TF between EOM and PZT loops, respectively before the replacement, Current UGF and cross-over frequency are 123.7 kHz and 14.3 kHz, respectively.

Then, we replaced the CMS from the original one to the modifed one. After that, we modifed the CMS setting and the related script in IMC guardian as follows.

  • K1:IMC-SERVO_SLOWFILTER: ON -> OFF 
  • K1:IMC-SERVO_SLOWPOL: - -> + (to compensate the sigh flip due to the turn off the slow filter)
  • K1:IMC-SERVO_IN1GAIN: 24 dB -> 4 dB (to compensate the gain increase of both slow and fast path by x10)

We confirmed IMC could be locked with this setting. However, since the offset was changed due to the CMS replacement, the normalized trans. power did not reached to ~1. Ushiba-san adjusted K1:IMC-SERVO_IN1GAIN_OFS_COMP_GAIN4_OFS not to change the value of K1:IMC-SERVO_MIXER_DAQ_OUT when IMC CMS IN1 opens and closes when IN1GAIN was 4 dB, and then tweaked K1:IMC-SERVO_OFS_COM_CALI_OFFSET to maximize IMC trans. power. (So you should perform the same adjustment if you want to change IN1GAIN from 4dB.) At any rate, IMC trans. power was maximized.

At last, we measured TFs after the replacement. Fig. 3 and Fig. 4 show the OLTF and the TF between EOM and PZT loops, respectively, after the replacement. There seems to be no significantly strange point in OLTF. On the other hand, there seems to be a small dip around the cross over frequency in the TF between EOM and PZT loops compared with before. However, the phase seems to be large enough around the cross over frequency. So they may be fine.

Replacement of the EOM amplifier from the x10 one to the x20 one

To enlarge the actuator range, we replaced an amplifier for a wideband EOM from x10 one (F10A, FLC electronics) to x20 one (made by Miyoki-san, klog35274). Miyoki-san told us the installation procedure of this amplifier. I followed the procedure.

  1. confirm that a dial on a regulator (DP1001), which is used as a 200V power supply, is 0. Then, connect a plug of the regulator to a power tap. 
    1. This time, I connected the plug to UPS in PSL
  2. set a current limit of the regulator to 60 mA.
  3. connect a plug for a fan in the amplifer to a power tap.
    1. This time, I connected the plug to the other power tap, not UPS.
  4. Turn on the regulator. And confirm 15 V are shown in a indicator on the regulator. Then, confirm that +/- 15 V is applied as an output voltage with a multimeter or something.
  5. At once, turn off the regulator before cabling
  6. Connect the regulator output port to a DC input port behind of the x20 amplifer box
  7. Connect the F30PV output port to the signal input port on the amplifer box
  8. Connect the signal output port on the box to EOM
  9. Turn on the regulator
  10. increase the applied voltage up to 190 V by rotating the dial on the regulator. (If the lamp on the upper right in the front panel of the regulator illuminated, reduce the voltage to 0 immediately. Then, contact with Miyoki-san)

Fig. 5 shows the amplifer box and Fig. 6 shows the regulator after the installation. Following this, we modified the CMS setting and the related script in IMC guardian. Because Miyoki-san's amplifer is an inversed amplifier, we flipped the sign of K1IMC-SERVO_FASTPOL from + to -. And we reduced K1:IMC-SERVO_FASTGAIN by 6 dB to compensate the gain increasing by x2 (=20/10) due to the replacement. After that, we confirmed IMC could be locked. Moreover, the red light did not illuminate during the lock acquisition and the lock.

Then, we measured the OLTF after the EOM amp. replacement. Fig. 7 shows the results. The gain seems to be not changed. On the other hand, the phase was rotated slightly by Miyoki-san's amplifier. However, since the phase margin is large enough at UGF, the loop is fine. (This time, UGF was not changed after the EOM amp. replacement so we did not measure the TF between EOM loop and PZT loops.)

---

IMC lock is keeping for 2 hours for now. We left IMC lock this night to check the stability.