Reports 1-1 of 1 Clear search Modify search
MIF (General)
takafumi.ushiba - 19:19 Tuesday 25 June 2024 (30110) Print this report
IRX spectrum characterization

Abstract:

I calculated the input equivalent noise of CARM CMS from the several DAQ channels when locking IRX, and compared them.
There are discrepancy between the channels, so some noisy components might exist in CARM CMS.

Detail:

One of the evaluations of characteristics of IRX LSC, I calculated the input equivalent noise of the CARM CMS from several .
Figure 1 shows the CARM medm screens when locking IRX.

1. K1:LSC-CARM_SERVO_MIXER_DAQ_OUT_DQ

K1:LSC-CARM_SERVO_MIXER_DAQ_OUT_DQ is calibrated to the voltage before the GENERIC FILTER (overall gain of 100).
So, calibration from K1:LSC-CARM_SERVO_MIXER_DAQ_OUT_DQ to CARM CMS input is just a gain of -10dB = 0.3162.

2. K1:LSC-CARM_SERVO_FAST_DAQ_OUT_DQ

K1:LSC-CARM_SERVO_FAST_DAQ_OUT_DQ is also calibrated to the voltage before the GENERIC filter (poles:100,100, zeros:10,10, DC gain:1), thanks to the filter modification today (klog30103).
So, calibration from K1:LSC-CARM_SERVO_FAST_DAQ_OUT_DQ to CARM CMS input is as follows:
poles: 0, 0, 1000, 1000
zeros: 5.3, 5.3, 10, 10
DC gain: 5.3**2 for compasating 5.3 Hz high pass filter, -11dB for FAST gain, 0.01**2 for compensating two BOOST filters, and -10dB for IN1 gain = 2.503e-4

3. K1:LSC-CARM_SERVO_SLOW_DAQ_OUT_DQ

K1:LSC-CARM_SERVO_SLOW_DAQ_OUT is also calibrated to the voltage before GENERIC filter (poles:100,100, zeros:10,10, DC gain:1).
So, calibration from K1:LSC-CARM_SERVO_SLOW_DAQ_OUT to CARM CMS input is as follows:
poles: 1000, 1000
zeros: 10, 10
DC gain: 0.01**2 for compensating two COM BOOST filters and -10dB for IN1 gain = 3.162e-5

4. K1:LSC-REFL_PDA1_RF45_I_ERR_DQ

K1:LSC-REFL_PDA1_RF45_I_ERR_DQ is calibrated to the input equivalent signals of analog whitening filters in the unit of DGS counts.
In addition, differential signals from I/Q demodulator to analog whitening filter is double of the signals sent to CARM CMS according to the circuit diagram (JGW-D1402413-v1).
So, calibration from K1:LSC-REFL_PDA1_RF45_I_ERR_DQ to CARM CMS input is as follows:
DC gain: 0.5 for differential to single end and 6.104e-4 for DGS counts to voltage = 3.052e-4

Figure 2 shows the CARM CMS input equivalent noise from each signal.
During the measurement, I engaged three dewhitening stages and 30dB overall gains not to be limited by ADC noise.

Discussion:

K1:LSC-CARM_SERVO_MIXER_DAQ_OUT_DQ is limited by ADC noise below several hundreds Hz but the other signals should not be limited by ADC noise.
At high frequency, all signals except for K1:LSC-REFL_PDA1_RF45_I_ERR_DQ are consistent, so they seems residual frequency noise.
Since K1:LSC-REFL_PDA1_RF45_I_ERR_DQ is roughly 2 times smaller than the others, I might miss a factor of 2 somewhere.

The spectrum of K1:LSC-REFL_PDA1_RF45_I_ERR_DQ is much higher than the others below 1kHz.
This implies that noise is induced after dividing error signals into two: the signals to CMS and the signal to analog whitening.
However, noise level seems too large if the noise comes from analog circuits, so something might happen in the I/Q demodulator or analog whitening filter.

The spectra of K1:LSC-CARM_SERVO_FAST_DAQ_OUT_DQ and K1:LSC-CARM_SERVO_SLOW_DAQ_OUT are similar but coherence between 20 Hz -to 1kHz is not close to 1.
ADC noise should be enough small, so some noisy component might exist after splitting feedback signals to fast and slow pathes.
Since FAST path signals are higher around several tens of Hz and lower than several hundreds of Hz, both FAST and SLOW pathes might have noise components.

Further investigation is necessary to characterize CARM CMS noise (noise budget of signle arm lock might be effective).

Note:

It would be better to do the same analysis for GRX and GRY PDH CMS.

Images attached to this report
Comments to this report:
hirotaka.yuzurihara - 10:45 Tuesday 09 July 2024 (30313) Print this report

Memo: the PATH to the xml file is /users/Commissioning/data/CARM/2024/0625/SPE_CARM_IRX_LOCKED_20240625.xml

hirotaka.yuzurihara - 15:54 Tuesday 09 July 2024 (30321) Print this report

I performed the crosscheck of the input quivelent noise of the input for the CARM CMS, when the IRX is locked (with Ushiba-san's help). I didn't find the critical mistake.

  • I checked the pole and zero of the generic filters (wiki page) used in CARM CMS.
  • I checked if the filter bank at the measurement time worked for the dewhitening filter and if the necessary FM switch was on.
  • I derived the gain factors from the DAQ values to input of the CARM CMS and confirmed they are correct.
  • I checked if the pole and zero for the calibtation from the DAQ to input of the CARM CMS is correct or not.
hirotaka.yuzurihara - 16:11 Tuesday 23 July 2024 (30526) Print this report

[Ushiba, Yuzu]
We performed the noise projection from the measured ADC noise (klog) to the each channel on the CARM CMS and made the plot of overlapping them. See the attached fig. All the projected ADC noise are located below the spectrum, except for 145 Hz (fig) and 375 Hz (fig) of LSC-CARM_SERVO_MIXER_DAQ

Details

  • The ADC noise were measured previously (klog) and was recorded in /users/DGS/measurements/ADC/K1ALS0/ALS0_ADC1_CH16-23.xml . I exported the data of CH[20, 21, 22] into the text. (/users/yuzu/work/20240709_CARM)
  • I imported them into the diaggui(/users/Commissioning/data/CARM/2024/0723/SPE_CARM_IRX_LOCKED_20240625.xm). I added the calibration to the imported data as follows.
  • ADC noise on LSC-CARM_SERVO_MIXER_DAQ
    • The switches for FM9 and FM10 were ON at the measurement of the spectrum.
    • The calibration from the ADC noise to the input equivalent noise of CARM CMS
      • Filter bank from the ADC noise to LSC-CARM_SERVO_MIXER_DAQ : gain 5e-3 * 0.0006104
      • Correction from LSC-CARM_SERVO_MIXER_DAQ to the input equivalent noise: 0.31622 (same as klog#30110)
      • Total gain: 9.6510344e-7
  • ADC noise on LSC-CARM_SERVO_FAST_DAQ
    • The switches for FM1 and FM10 were ON at the measurement of the spectrum.
    • The calibration from the ADC noise to the input equivalent noise of CARM CMS
      • Filter bank from the ADC noise to LSC-CARM_SERVO_FAST_DAQ : gain 0.0006104*0.5, filter zpk([100, 100], [10, 10])
      • Correction from LSC-CARM_SERVO_FAST_DAQ to the input equivalent noise: gain 0.0002503, filter zpk([5.3, 5.3, 10, 10], [0, 0, 1000, 1000]) (same as klog#30110)
      • Total gain: 7.639156e-8
  • ADC noise on LSC-CARM_SERVO_SLOW_DAQ
    • The switches for FM1 and FM10 were ON at the measurement of the spectrum.
    • The calibration from the ADC noise to the input equivalent noise of CARM CMS
      • Filter bank from the ADC noise to LSC-CARM_SERVO_FAST_DAQ : gain 0.0006104*0.5, filter zpk([100, 100], [10, 10])
      • correction from LSC-CARM_SERVO_FAST_DAQ to the input equivalent noise: gain 3.162e-5, filter zpk([10, 10], [1000, 1000]) (same as klog#30110)
      • Total gain: 9.650424e-9
Images attached to this comment
takahiro.yamamoto - 12:39 Wednesday 24 July 2024 (30544) Print this report
Do you know where does the difference on RF45_I_ERR between the cases of IRX (klog#30526) and IRY (klog#30529) come from?

According to the other spectra (MIXER_, FAST_, and SLOW_DAQ), the noise which limits RF45_I_ERR seems to exists on the pick off path to the digital system not a real signal in the control loop. If it really exists in the control loop, it should be able to be seen also on MIXER_, FAST_, and SLOW_DAQ which are more sensitive channel on that frequency region. On the other hand, if it comes from simple circuits and/or cabling problems on the pick off path to the digital system, it should be almost same in the both cases of IRX and IRY.
takafumi.ushiba - 23:59 Wednesday 24 July 2024 (30558) Print this report

In fact, we are not so sure the reason and this is one of the mysteries we need to investigate.

Search Help
×

Warning

×