[Hirata, Akutsu-san, Dan Chen-san, Ikeda-san]
1. Measurement of Breadboad height
We measured breadboard height before fixing breadboard. The reference point is on the pillar at -Y +X side of the chamber. The result is attached.
2. Fix the breadboard
We lower 4 foot screw of breadboard, and fix the breadboard.
3. Recovery work
We started recovering work.
- We selected Guardian status "SAFE".
- First, we checked OSEM flag positions by eye.
- We centered tilt Oplev by moving IM moving masses, and IM yaw. Sometimes we checked OSEM flags and moved OSEM body until it doesn't hit.
- Some OSEM position is still not centered. We should adjust later. Signals are between 2000 and 8000.
- Next, we tried to adjust length Oplev by Traverser, but it looks no change. We talked with Aso-san, and realized Seg1 and Seg 4 were satulated. We need to move more.
- Ushiba-san advised Traverser should be move to AR side direction.
Tomorrow, we will continue to adjust length Oplev using Traverser. After that we will adjust tilt Oplev and check OSEM position again.