[Komori, Michimura]
During 2025, arm cavity finesse dropped from ~1400 to ~1300 for Xarm and ~1300 to ~1200 for Yarm.
This can be explained by arm cavity round-trip loss increase from ~50ppm to ~600ppm.
This naively means that the arm cavity loss when locked was as high as ~40%, assuming that ITM reflectivities are constant.
This is not compatible with observed power recycling gain drop, which suggests that ITM reflectivities have also dropped during 2025.
Still, it is hard to explain both power recycling gain drop and arm cavity finesse drop simultaneously.
Assumptions:
- Mirror reflectivities are well measured as follows, and are constant over 2025.
T['PRM']=10.35/100 # https://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/private/DocDB/ShowDocument?docid=5744 (uncetainties unknown)
T['ITMX']=ufloat(0.444/100,1.76e-5) # PhysRevApplied.14.014021
T['ETMX']=ufloat(6.8e-6,3.87e-7) # PhysRevApplied.14.014021
T['ITMY']=ufloat(0.479/100,2.17e-5) # PhysRevApplied.14.014021
T['ETMY']=ufloat(6.92e-6,5.16e-6) # PhysRevApplied.14.014021
What we did:
- Fetched the minute trenda data of the following channel using ndscope from Mar 1, 2025 to Dec 12, 2025 0:00 JST, and exported into hdf5 file.
K1:GRD-LSC_LOCK_STATE_N
K1:LAS-POW_IMC_DC_INMON
K1:LSC-POP_PDA2_RF90_I_ERR_DQ
K1:LSC-TR_IRX_NORM_OUT_DQ
K1:LSC-TR_IRY_NORM_OUT_DQ
- Plotted the data if LSC_LOCK_STATE_N>=9998 (WORKING_WITH_OBSERVATION_WITHOUT_LINES). See left panels of the attachment.
- Also plotted measured finesses by loading the following files. See right top panel.
/users/Commissioning/data/Finesse/Xarm/summary.txt
/users/Commissioning/data/Finesse/Yarm/summary.txt
- Did a similar analysis done by Takidera-kun (see klog #34857). In this analysis, we first derived the arm cavity round-trip loss (RTL) from the measured finesse using the following equations. Actually, using the raw measured finesse values, estimated RTL will be less than 0. This could be due to fitting errors of the finesse measurement code (to be fixed by Oshino-san). To make the round-trip loss values compatible with klog #30823 at nominal finesse days, measured finesse were intentionally corrected for 40 for X (which is roughly consistent with Oshino-san's observations) and 30 for Y, and added extra +/- 10 uncertainties (See right top and middle panels).
Finesse = pi*sqrt(rITM*rloss)/(1-rITM*rloss)
ArmRTL = 1-rloss**2
- From normalized arm cavity transmission, losses in the PRC, including losses from the arms (RTL enhanced by arm gain) can be estimated using the following equations. See blue and red dots in the right bottom panel.
TR_IRX_NORM = PRG / T_PRM = 1/(1-rPRM*rfp)**2
Tprc = 1-rfp**2
- We can also estimate the losses from the arms using the following equation. See cyan and magenta dots in the right bottom panel. To my surprise, this was not compatible with the estimated Tprc.
Tarm = 1 - (-rITM+tITM**2*rloss/(1-rITM*rloss))**2
Next:
- Consider that ITM reflectivities are not constant. Somehow need to be estimated from sideband power recycling gain trend (but this is degenerated with losses in PRC). Also, drop in the ITM reflectivities alone cannot explain almost 1/3 drop in the PRG.
- Attribute all other causes of arm cavity finesse drop to arm cavity round-trip loss increase, and see if this senario can explain carrier power recycling gain measurements.