Reports 1-1 of 1 Clear search Modify search
VIS (SRM)
yoshinori.fujii - 18:09 Thursday 21 February 2019 (8177) Print this report
Comment to Installation work (7859)

After pre-amp board replacement, the geophones were calibrated again and the obtained paramters are the following:

geo_name Ge [V/(m/s)] f0 [Hz] eta
geo-H1 243.96 1.0393 0.3220
geo-H2 240.33 1.0209 0.3263
geo-H3 240.41 1.1301 0.3011

The frequency responce of the geophone in V/(m/s) was estimated with the seismometer signal at IXV, like raw_geophone_sig / seismometer.
 

Images attached to this comment