[Izumi, Eleonora, Yokogawa]
Related klog: #6239
Today, the new PROMETHEUS laser for ALS (Y arm) system was delivered, so we measured frequency of this laser by tuning the crystal temperature.
The attached figure shows the result. Green dots show the measurement result of the new PROMETHEUS laser. Also, the measurement result of the main laser and the PROMETHEUS laser for X arm measured on Sep. 21st are shown in red and blue dots respectively. The frequency differences between the main laser and two PROMETHEUS lasers need to be around 40 MHz since the beat-note signal is locked to 40 MHz by PLL. However, the overlap range (gray region) is very narrow and the frequency of the main laser can be set only every 1.5 GHz in order to lock the main laser to the reference cavity. (FSR of the reference cavity is 1.5 GHz.)
I believe that the beat-note signal for ALS Y arm system can be seen...