Reports 1-1 of 1 Clear search Modify search
ISC (IMC)
kentaro.komori - 11:28 Friday 30 October 2026 (37773) Print this report
IMC length measurement after tuning

[Ushiba, Yokozawa, Kawakami, Komori]

Abstract:

We measured the IMC round-trip length and its free spectral range (FSR) as follows:

L_imc = 53.281939(1) m
f_m = 5.6265306(1) MHz

The actual MCE displacement was estimated to be 10.256(1) mm, in excellent agreement with the planned displacement of 10.26 mm.

Details:

After moving the MCE, we measured the IMC length and FSR again to determine the actual MCE displacement.
The measurement method and setup were exactly the same as those used in the previous measurements, in particular the most recent measurement performed immediately before the MCE adjustment (klog:37657).

First, we set the seed sideband frequency to the target value chosen to optimize the separation between the RF sidebands and the arm higher-order modes (HOMs), which was the main purpose of the MCE displacement.
The peaks at the modulation frequency in the IFO REFL PDH signals are shown by the green and brown traces in Fig. 1.
Next, we tuned the sideband frequency and found that the peak height was minimized at 5.6265306(1) MHz, as shown by the red and blue traces.
This corresponds to an IMC round-trip length of 53.281939(1) m.

We had planned to move the MCE by 10.26 mm to achieve the optimal RF-sideband/HOM separation, while the displacement during the actual work was estimated to be 10.22(5) mm (klog:37718).
From the measured change in the IMC round-trip length, we can now estimate the actual MCE displacement more accurately to be 10.256(1) mm, which is in excellent agreement with the planned value.

We also observed mixing between the I- and Q-phase components in the demodulated signals.
Therefore, the demodulation phase should be retuned accordingly.
In addition, if we want to adjust the offset of the IMC length error signal, the error-signal offset and the sideband frequency should be optimized together in a future measurement.