[Fujimoto, Komori]
Abstract:
We measured the IMC FSR to be 5.6243653(1) MHz, corresponding to an IMC round-trip length of 53.302451(1) m.
To reach the target RF sideband frequency, we will move the MCE by 10.26 mm toward the MCI/MCO.
Details:
We tuned the frequency of the seed RF sideband used for the IFO LSC so as to minimize the peak produced by laser-frequency excitation, using the same method as described in klog:23115.
The optimal frequency was found to be 5.6243653(1) MHz, corresponding to an IMC round-trip length of 53.302451(1) m.
This value differs by 13 µm from the previous measurement performed in 2024 (klog:29385).
The target sideband frequency is 5.6265328 MHz.
Therefore, the IMC round-trip length should be shortened by 20.53 mm.
Since the MCE displacement changes the round-trip length by twice the mirror displacement, we will move the MCE by 10.26 mm toward the MCI/MCO.