ISC (IMC)
Seidai Kawakami - 9:35 Thursday 22 October 2026 (37718) Print this report
Comment to Results of MCE adjustment (37713)

There are two corrections to this report.

  1. The uncertainty in the displacement was initially reported as 0.005 mm. However, the readings on the right side showed a scatter of approximately 0.1 mm, so an uncertainty of 0.1 mm was assigned to the right-side measurement. Since the final displacement was calculated by averaging the left- and right-side displacements, the resulting uncertainty was 0.05 mm.

  2. In the measurement table, the Remarks entry for the row with a left-side distance of 13.95 mm and a right-side distance of 16.70 mm should read “Removed the clamp” instead of “Removed the pusher.”