Reports 1-1 of 1 Clear search Modify search
MIF (General)
shun.saito - 22:14 Thursday 25 June 2026 (37123) Print this report
Attempt to observe SRC flash while operating the PLL

[Tanaka, Fujimoto, Saito]

When the LO frequency was adjusted relative to the beat frequency so that the error signal became close to 0 V, it was found that the beat frequency and LO frequency matched at higher frequencies, whereas an offset appeared at lower frequencies. Next, in order to achieve locking using only the SR560, the SR560 was configured as a first-order low-pass filter with a cutoff frequency of 1 Hz and a gain of 200. Under this condition, locking was successfully achieved. Measurement of the open-loop transfer function showed that the UGF was approximately 17 Hz. Next, we attempted to use the FM modulation function of the LO source (Keysight E8663D) in order to sweep the LO frequency and scan the SRY. However, the FM modulation option was not installed. Therefore, the function generator of the Moku:Lab was used instead. When the frequency modulation was performed at a rate of 1 Hz, no SRY flashes were observed on the OMC REFL PD. This is likely because the frequency modulation was carried out at 1 Hz, whereas the cutoff frequency of the high-pass filter in the SR560 used for the OMC REFL PD signal was 300 Hz. Therefore, the cutoff frequency was lowered to 100 Hz, but the main-laser noise became comparable to the amplitude of the sub-laser flashes, so the cutoff frequency was restored to 300 Hz. In addition, increasing the frequency modulation rate of the LO caused the beat signal waveform to become distorted.
 

  • First, the LO frequency was adjusted relative to the beat frequency so that the error signal became close to 0 V. The resulting frequencies were as follows:

    Beat frequency    LO frequency
    33 MHz                    54 MHz
    78 MHz                    85 MHz
    139 MHz                 139 MHz
    179 MHz                 179 MHz

    Therefore, at higher frequencies there appears to be no offset, and the beat frequency matches the LO frequency. Furthermore, at lower frequencies, the beat signal and LO signal were directly observed using the Moku:Lab oscilloscope. The measured beat frequency agreed with the frequency observed on the Moku:Lab spectrum analyzer after the signal had been split by the power splitter, and the LO frequency agreed with the set value. Therefore, the observed offset appears to originate from the PFD.
     

  • Next, a DC voltage was applied to the SR560 to determine its allowable input and output voltage ranges before overload occurred. The maximum allowable input voltage was found to be 2 V, while the maximum output voltage was 5 V. When the error signal was near 0 V, observation of the error signal with a 100 kHz low-pass filter showed fluctuations of approximately 4 Vpp. Therefore, a 20 dB attenuator was inserted to ensure that the signal could be safely input to the SR560. Furthermore, when the error signal was 150 mV, the beat frequency and LO frequency differed by approximately 10 MHz, corresponding to a sensing efficiency of approximately 15 nV/Hz. To achieve locking using only the SR560, the SR560 was configured as a first-order low-pass filter with a cutoff frequency of 1 Hz and a gain of 200, and stable locking was achieved. The current control loop from the RFPD output to the sub-laser PZT is as follows:

    RFPD→ 12 MHz high-pass filter→ 20 dB RF amplifier→ 45 dB RF amplifier→ 10 dB attenuator→ PFD→ 20 dB attenuator→ 100 kHz low-pass filter→ SR560 (1 Hz cutoff frequency, gain ×200, first-order low-pass filter)→ Sub-laser PZT

    The open-loop transfer function was then measured, yielding a UGF of approximately 17 Hz (Fig. 1).
     

  • Next, in order to scan the SRY by sweeping the LO frequency, we attempted to use the FM modulation function of the LO source (Keysight E8663D). However, this option was not installed. Therefore, a 1 Vpp sinusoidal signal was generated using the Moku:Lab function generator and used as the LO signal. The LO frequency was first matched to the beat frequency and then frequency-modulated by ±2 MHz at a modulation rate of 1 Hz (Fig. 2). However, no SRY flashes were observed on the OMC REFL PD. This is likely because the modulation rate of 1 Hz is below the 300 Hz cutoff frequency of the high-pass filter in the SR560 used for the OMC REFL PD signal. The cutoff frequency was therefore reduced to 100 Hz, but the main-laser noise became comparable in magnitude to the sub-laser flash signal, so the cutoff frequency was returned to 300 Hz. The dominant main-laser noise frequency was approximately 60 Hz. In addition, the frequency modulation rate of the LO was increased, but the shape of the beat signal became distorted.

Images attached to this report
Comments to this report:
Hiroki Fujimoto - 2:47 Friday 26 June 2026 (37124) Print this report

[Saito, Hirose, Tanaka, Fujimoto]

Abstract

We found that the intensity noise observed with the OMC REFL PD, which will be used for the SRCL/PRCL measurement, can be improved by aligning SRY.
This result suggests that we may be able to lower the cut-off frequency of the SR560 (high-pass) connected to the OMC REFL PD output from the current value of 300 Hz.
This may allow us to reduce the scan speed of the auxiliary laser.

Details

Improvement of the OMC REFL intensity noise by SRY alignment

In the evening, we measured the intensity fluctuation of the OMC REFL PD while the main laser was locked to SRY.
We found that the spectrum had become worse than the one measured yesterday.

In Fig. 1, the blue trace shows the spectrum measured in the evening, and the green trace shows the spectrum measured yesterday.
In particular, the 60 Hz line noise became prominent.

When we lowered the cutoff frequency of the SR560 (high-pass) attached to the OMC REFL PD from 300 Hz to 100 Hz today, a noise comparable in size to the auxiliary laser flashes appeared.
This noise was likely coming from the 60 Hz line noise, since its frequency was roughly around 70 Hz when we checked on the oscilloscope.

We then realigned SRY so as to maximize the AS DC power with ADS and manual alignment of ITMY.
As a result, the OMC REFL spectrum was reduced, as shown by the brown trace in Fig. 1, and became quieter than yesterday’s measurement.

In addition, the 60 Hz power-line noise was significantly reduced, and the noise appears to have moved to harmonics such as 120 Hz.

Fig. 2 shows the evolution of the OMC REFL spectrum during the process of improving the alignment. It can be seen that the intensity noise was gradually reduced.


Investigation of the origin of the 60 Hz power-line noise

Since the 60 Hz power-line noise was improved by the SRY alignment, mainly in the pitch direction, and the component moved to harmonics, it is suspected that one of the suspensions related to SRY is moving at 60 Hz in pitch.

Therefore, I measured the pitch and yaw spectra of the related suspensions and looked for suspensions showing a 60 Hz line.

The result is shown in Fig. 3. A 60 Hz component can be seen in the pitch spectrum of SR3. Since this component became invisible in the OMC REFL signal after improving the SRY pitch alignment, SR3 pitch motion is likely the cause of the 60 Hz noise seen in OMC REFL.

Plans for tomorrow

We plan to reduce the high-pass cutoff frequency of the SR560 attached to the OMC REFL PD as much as possible while keeping the OMC REFL intensity noise low by using the improved SRY alignment. This is expected to allow us to reduce the scan speed of the auxiliary laser.

And the current UGF of the PLL is about 20 Hz. We will try to increase the control bandwidth, for example by removing the 20 dB attenuator placed after the PFD, so that the auxiliary laser can follow a faster LO scan.

Images attached to this comment
shun.saito - 14:23 Sunday 28 June 2026 (37130) Print this report

[Smith, Hirose, Fujimoto, Saito]

The 20 dB attenuator placed before the SR560 input was removed, and the SR560 gain was increased to 2000. In addition, a Moku:Lab was inserted before the SR560, and an integrator with a cutoff frequency of 100 Hz was added. Measurement of the open-loop transfer function showed that the UGF was approximately 2 kHz. With the cutoff frequency of the high-pass filter in the SR560 used for the OMC REFL PD signal set to 300 Hz, SRY flashes were observed when the LO frequency was modulated by ±2 MHz at a modulation rate of 300 Hz. Furthermore, SRY flashes were also observed when the cutoff frequency of the SR560 high-pass filter was changed to 100 Hz and the modulation rate was reduced to 100 Hz. However, the noise level was higher than in the 300 Hz cutoff-frequency case.
 

  • First, as in the previous experiment (klog:37123), the PLL was locked using only the SR560 by configuring it as a first-order low-pass filter with a cutoff frequency of 1 Hz and a gain of 200. The open-loop transfer function was then measured, yielding a UGF of approximately 2.5 Hz (Fig. 1). The reason why the UGF decreased compared with the previous result (klog:37123) is unclear. Next, the 20 dB attenuator before the SR560 input was removed, and the SR560 gain was increased to 2000. Measurement of the open-loop transfer function under these conditions showed a UGF of approximately 2 kHz. The corresponding open-loop transfer function is shown by the light orange trace in Fig. 2. Subsequently, a Moku:Lab was inserted before the SR560, and an integrator with a cutoff frequency of 100 Hz was added. The measured open-loop transfer function again showed a UGF of approximately 2 kHz. The corresponding transfer function is shown by the orange trace in Fig. 2.
     
  • Next, while attempting to scan the SRY, it was noticed that the DC signal from the OMC REFL PD was approximately -300 counts. When the light incident on the PD was blocked using a sensor card, the signal returned to approximately 0 counts (Fig. 3). The reason for the offset of approximately -300 counts remains unknown.
     
  • Next, a 1 Vpp sinusoidal signal generated by the Moku:Lab function generator was used to modulate the LO frequency. The LO frequency was first set equal to the beat frequency and then frequency-modulated by ±2 MHz at a modulation rate of 300 Hz. Under these conditions, with the cutoff frequency of the SR560 high-pass filter used for the OMC REFL PD signal set to 300 Hz, four peaks were observed within one 300 Hz modulation cycle (Fig. 4). Since the FSR of the cavity is approximately 2.2 MHz, these peaks are believed to correspond to SRY flashes. Furthermore, when the modulation rate was reduced to 100 Hz and the cutoff frequency of the SR560 high-pass filter was also reduced to 100 Hz, four peaks were again observed within one 100 Hz modulation cycle (Fig. 5). However, the noise level was higher than in the case with the 300 Hz cutoff frequency.
Images attached to this comment
yoichi.aso - 17:25 Monday 29 June 2026 (37135) Print this report

I found a frequency counter with the frequency range up to 200MHz in NAOJ.

It will be shipped to Kamioka tomorrow.

 

shun.saito - 20:46 Monday 29 June 2026 (37138) Print this report

[Smith, Fujimoto, Saito]

The cutoff frequency of the high-pass filter in the SR560 used for the OMC REFL PD signal was set to 300 Hz. As in the previous experiment (klog:37130), the PLL was locked using the SR560 configured as a first-order low-pass filter with a cutoff frequency of 1 Hz and a gain of 2000, together with an additional integrator with a cutoff frequency of 100 Hz implemented in Moku:Lab. The LO frequency was frequency-modulated with a sensitivity of ±10 MHz/V. When a 300 Hz, 800 mVpp sinusoidal or triangular signal generated by a function generator was used as the external modulation signal, SRY flashes were successfully observed. Furthermore, data for an SRY scan were acquired using a frequency modulation sensitivity of ±15 MHz/V. The FSR and the length of the SRY will be determined from these data.
 

  • Since the main-laser noise was smaller when the cutoff frequency of the SR560 high-pass filter used for the OMC REFL PD signal was set to 300 Hz than when it was set to 100 Hz, the cutoff frequency was set to 300 Hz.
     
  • Next, only the sub-laser beam was injected, and the alignment was adjusted to maximize the fringe amplitude. The main-laser beam was then injected as well. The alignment of the PLL path was also optimized to maximize the beat signal amplitude.
     
  • As in the previous experiment (klog:37130), the PLL was locked using the SR560 configured as a first-order low-pass filter with a cutoff frequency of 1 Hz and a gain of 2000, together with an additional integrator with a cutoff frequency of 100 Hz implemented in Moku:Lab. A 1 Vpp sinusoidal signal generated by the Moku:Lab function generator was used as the LO signal. The LO frequency was adjusted to approximately match the beat frequency and then finely tuned so that the feedback signal was close to 0 V. The LO frequency was then frequency-modulated with a sensitivity of ±10 MHz/V. When a 300 Hz, 800 mVpp sinusoidal signal generated by a function generator was used as the external modulation signal, the result shown in Fig. 1 was obtained. The upper plot shows the OMC REFL PD signal, while the lower plot shows the external modulation signal used for the frequency modulation. When the external signal was changed to a triangular waveform, the result shown in Fig. 2 was obtained. In both Fig. 1 and Fig. 2, the sub-laser PZT was swept over a range of ±4 MHz, and approximately seven peaks were observed during one period of the external modulation signal. Since the FSR is approximately 2.2 MHz, these peaks are believed to correspond to SRY flashes. Increasing the frequency of the external modulation signal resulted in cleaner peak shapes, but the number of observed peaks decreased. Therefore, a modulation frequency of approximately 300 Hz appears to be optimal.
     
  • Finally, an SRY scan was performed using a frequency modulation sensitivity of ±15 MHz/V, and the OMC REFL PD signal together with the external modulation signal were recorded. The FSR and the corresponding SRY length will be determined from these data.
Images attached to this comment
shun.saito - 3:29 Wednesday 01 July 2026 (37145) Print this report

The SRY scan data acquired in the previous experiment (klog:37138) were divided into segments according to the time intervals between changes in the slope of the triangular waveform used for frequency modulation, and each segment was fitted individually. When the same initial values for the fitting parameters were used for all segments, some of the fits failed to converge properly. However, these problematic cases could be fitted successfully by changing the initial parameter values, suggesting that individual initial values should be assigned for each data segment. Using identical initial values for all segments, the resulting SRY length was found to be 64.94 ± 0.02 m.
 

  • First, the OMC REFL PD signal data acquired in the previous experiment (klog:37138) were normalized. The signal used for frequency modulation was converted from counts to frequency using the conversion factors of 610 μV/count and 15 MHz/V. Next, the data were divided according to the intervals between changes in the slope of the triangular waveform used for frequency modulation, and each segment was fitted using scipy.optimize.curve_fit. The fitting function was

    Φ=4πL(f+f_offset)/c
    P_t=A/(1+B(sin(Φ/2))^2)

    where f is the modulated frequency and c is the speed of light. The fitting parameters were L, f_offset, A, and B. The initial values of the fitting parameters were set as follows:

    L=65 m (the expected SRY length),
    f_offset=282 THz (the laser frequency),
    A = the maximum value of the OMC REFL PD signal within the fitting interval,
    B=6.48, corresponding to 4×finesse/π^2 for an assumed SRY finesse of 4.

    In addition, near the times at which the slope of the triangular modulation waveform changed, the frequency derived from the recorded data might differ from the actual frequency variation. Therefore, for each interval between successive slope changes, the first one-sixth and the last one-sixth of the data were excluded from the fitting procedure. The fit corresponding to the smallest uncertainty in the SRY length is shown in Fig. 1, while the fit corresponding to the largest uncertainty is shown in Fig. 2. The fit in Fig. 1 appears satisfactory, whereas the fit in Fig. 2 appears poor. Although the fit in Fig. 2 can be improved by changing the initial parameter values, doing so causes some other datasets to fit poorly. Therefore, it is likely necessary to choose the initial parameter values individually for each dataset. The SRY lengths obtained from all fits are shown in Fig. 3. Here, up sweep refers to the intervals in which the slope of the triangular modulation waveform was positive, while down sweep refers to the intervals in which the slope was negative. The mean values and standard errors calculated from all fitted SRY lengths are:

    Up sweep: 64.89±0.03 m
    Down sweep: 64.99±0.02 m
    →Overall: 64.94±0.02 m