Reports 1-1 of 1 Clear search Modify search
MIF (General)
shun.saito - 20:46 Monday 29 June 2026 (37138) Print this report
Comment to Attempt to observe SRC flash while operating the PLL (37123)

[Smith, Fujimoto, Saito]

The cutoff frequency of the high-pass filter in the SR560 used for the OMC REFL PD signal was set to 300 Hz. As in the previous experiment (klog:37130), the PLL was locked using the SR560 configured as a first-order low-pass filter with a cutoff frequency of 1 Hz and a gain of 2000, together with an additional integrator with a cutoff frequency of 100 Hz implemented in Moku:Lab. The LO frequency was frequency-modulated with a sensitivity of ±10 MHz/V. When a 300 Hz, 800 mVpp sinusoidal or triangular signal generated by a function generator was used as the external modulation signal, SRY flashes were successfully observed. Furthermore, data for an SRY scan were acquired using a frequency modulation sensitivity of ±15 MHz/V. The FSR and the length of the SRY will be determined from these data.
 

  • Since the main-laser noise was smaller when the cutoff frequency of the SR560 high-pass filter used for the OMC REFL PD signal was set to 300 Hz than when it was set to 100 Hz, the cutoff frequency was set to 300 Hz.
     
  • Next, only the sub-laser beam was injected, and the alignment was adjusted to maximize the fringe amplitude. The main-laser beam was then injected as well. The alignment of the PLL path was also optimized to maximize the beat signal amplitude.
     
  • As in the previous experiment (klog:37130), the PLL was locked using the SR560 configured as a first-order low-pass filter with a cutoff frequency of 1 Hz and a gain of 2000, together with an additional integrator with a cutoff frequency of 100 Hz implemented in Moku:Lab. A 1 Vpp sinusoidal signal generated by the Moku:Lab function generator was used as the LO signal. The LO frequency was adjusted to approximately match the beat frequency and then finely tuned so that the feedback signal was close to 0 V. The LO frequency was then frequency-modulated with a sensitivity of ±10 MHz/V. When a 300 Hz, 800 mVpp sinusoidal signal generated by a function generator was used as the external modulation signal, the result shown in Fig. 1 was obtained. The upper plot shows the OMC REFL PD signal, while the lower plot shows the external modulation signal used for the frequency modulation. When the external signal was changed to a triangular waveform, the result shown in Fig. 2 was obtained. In both Fig. 1 and Fig. 2, the sub-laser PZT was swept over a range of ±4 MHz, and approximately seven peaks were observed during one period of the external modulation signal. Since the FSR is approximately 2.2 MHz, these peaks are believed to correspond to SRY flashes. Increasing the frequency of the external modulation signal resulted in cleaner peak shapes, but the number of observed peaks decreased. Therefore, a modulation frequency of approximately 300 Hz appears to be optimal.
     
  • Finally, an SRY scan was performed using a frequency modulation sensitivity of ±15 MHz/V, and the OMC REFL PD signal together with the external modulation signal were recorded. The FSR and the corresponding SRY length will be determined from these data.
Images attached to this comment
Search Help
×

Warning

×