Shingo Hido - 16:52 Monday 30 June 2025 (34412) Print this report
Full calibration measurements

with Dan, YamaT, SawadaT

We performed full measurements for the IFO calibration.
The analysis results will be posted at a later date.

Notes:

The DARM-related measurements were conducted after adjusting the feedforward gains (see klog#34410).
Following the same method as klog#34040 and klog#34033, we performed measurements to observe the changes in DC caused by differences in the number of whitening filter stages for ITMX-TM and BS-TM.

Measurement files (`/users/CAL/current/measurements/2025/0603/0920_*`):
0920_BS_TM_PRCL.xml   : Whitening = 1 stage
0920_BS_TM_PRCL-1.xml  : Whitening = 0 stage
0920_ITMX_TM_DARM_PRFPMI_WFOFF.xml : Whitening = 0 stage
0920_ITMX_TM_DARM_PRFPMI_WFON.xml  : Whitening = 3 stages

Comments to this report:
takahiro.yamamoto - 1:50 Tuesday 01 July 2025 (34416) Print this report
Following table is a list of estimated calibration parameters.
  • Each value is estimated based on JGW-L2314903.
  • Estimated values by Pcal and Free-swing are consistent with each other at approximately 3%.
  • Previous full measurement which is the reference of comparison is klog#34033.
  • There is no significant change in all parameters.
  • Numbers in parentheses indicate a number of engaged de-whitening filters.
    • Low-frequency (<1Hz) zero/pole in disengaged de-whitening filters (see also klog#33874) are compensated based on circuit measurement only for ETMX in klog#34229 and klog#34244.
    • So this effect must be considered for other suspensions.
 
  Free-swing XPcal Diff. from prev. Figures
H_mich 1.115e+09 +/- 0.0067e+09 N/A -2.8% Fig.1
A_BS(0) 6.180e-11 +/- 0.039e-13 6.000e-11 +/- 0.048e-11 +0.8% (Free-swing) Fig.2-3 (for Free-swing)
A_ITMX(0)/A_BS(0) 0.04956 +/- 0.00038 same as left -0.2% Fig.4-5
A_ITMX(0) 3.063e-12 +/- 0.030e-12 3.3583e-12 +/- 0.0043e-12 +0.7% (Free-swing)
A_ITMX(3)/A_ITMX(0) 1.1292 +/- 0.0018 same as left +0.1% Fig.6-7
A_ITMX(3) 3.459e-12 +/- 0.035e-12 3.3583e-12 +/- 0.0043e-12 +0.7% (Free-swing)
A_ETMX_TM/A_ITMX(3) 1.1456 +/- 0.0011 same as left -7.9% Fig.8-9
A_ETMX_TM 3.962e-12 +/- 0.040e-12 3.8574e-12 +/- 0.0033e-12 -0.6% (XPcal) Fig.10-11 (for XPcal)
A_ETMX_IM/A_ETMX_TM 42.1 +/- 2.4 N/A -7.8% Fig.12-13
A_ETMX_IM 1.670e-10 +/- 0.097e-10 1.68e-10 +/- 0.10 -8.7% (XPcal) Fig.14-15 (for XPcal)
H_DARM 2.211e+12 +/- 0.023e+12 2.2775e+12 +/- 0.0071e+12 +0.9% (XPcal) Fig.16-19
1/H_DARM 4.522e-13 +/- 0.048e-13 4.391e-13 +/- 0.014e-13 N/A
A_BS(1)/A_BS(0) 1.0423 +/- 0.0048 same as left +0.5% (Free-swing) Fig.20-21
A_BS(1) 6.441e-11 +/- 0.050e-11 6.254e-11 +/- 0.057e-11 +1.3% (Free-swing)
H_MICH 4.628e+08 +/- 0.036e+08 4.767e+08 +/- 0.044e+08 +9.8% (Free-swing) Fig.22-25
1/H_MICH 2.161e-09 +/- 0.017e-09 2.098e-09 +/- 0.019e-09 N/A
A_PRM(1)/A_BS(1) 27.21 +/- 0.12 same as left +1.0% Fig.26-27
M2P FF gain 0.03675 +/- 0.00016 same as left N/A
A_PRM 8.764e-10 +/- 0.078e-10 8.509e-10 +/- 0.086e-10 +2.3% (Free-swing)
H_PRCL 2.057e+09 +/- 0.018e+09 2.118e+09 +/- 0.021e+09 +4.3% (Free-swing) Fig.28-31
1/H_PRCL 4.862e-10 +/- 0.043e-10 4.720e-10 +/- 0.048e-10 N/A
A_ITMX(3)/A_BS(1) 0.05369 +/- 0.00049 same as left N/A
M2D FF gain 18.6 +/- 3.2 same as left N/A

Images attached to this comment
takahiro.yamamoto - 13:14 Tuesday 01 July 2025 (34423) Print this report
Some mistakes were fixed.
1) Wrong measurement file was used for DARM optical gain estimation.
I had used /users/CAL/current/measurements/2025/0630/0920_DARM_OLG.xml in which the measurement was interrupted.
Correct (re-measured) one is /users/CAL/current/measurements/2025/0630/0920_DARM_OLG-1.xml.
I re-analyzed with correct measurement results and H_DARM and 1/H_DARM were updated.
Finally, the difference from previous results is ~3% and a trend of difference will be reported by Hido-kun.

2) Comparison of the actuator efficiency ratio of ETMX and ITMX was done in wrong configuration.
I had compared A_ETMX_TM/A_ITMX_TM(3) in this time and A_ETMX_TM/A_ITMX_TM(0) in the previous measurement.
So I re-evaluated the difference by using A_ETMX_TM/A_ITMX_TM(3) in the previous measurement.

And also comparison between YPcal and XPcal was added in the table.
  Free-swing XPcal Diff. from prev. Figures
H_mich 1.115e+09 +/- 0.0067e+09 N/A -2.8% Fig.1
A_BS(0) 6.180e-11 +/- 0.039e-13 6.000e-11 +/- 0.048e-11 +0.8% (Free-swing) Fig.2-3 (for Free-swing)
A_ITMX(0)/A_BS(0) 0.04956 +/- 0.00038 same as left -0.2% Fig.4-5
A_ITMX(0) 3.063e-12 +/- 0.030e-12 3.3583e-12 +/- 0.0043e-12 +0.7% (Free-swing)
A_ITMX(3)/A_ITMX(0) 1.1292 +/- 0.0018 same as left +0.1% Fig.6-7
A_ITMX(3) 3.459e-12 +/- 0.035e-12 3.3583e-12 +/- 0.0043e-12 +0.7% (Free-swing)
A_ETMX_TM/A_ITMX(3) 1.1456 +/- 0.0011 same as left -7.9%
-0.9%
Fig.8-9
A_ETMX_TM 3.962e-12 +/- 0.040e-12 3.8574e-12 +/- 0.0033e-12 -0.6% (XPcal) Fig.10-11 (for XPcal)
A_ETMX_IM/A_ETMX_TM 42.1 +/- 2.4 N/A -7.8% Fig.12-13
A_ETMX_IM 1.670e-10 +/- 0.097e-10 1.68e-10 +/- 0.10 -8.7% (XPcal) Fig.14-15 (for XPcal)
H_DARM 2.211e+12 +/- 0.023e+12
2.1246e+12 +/- 0.023e+12
2.2775e+12 +/- 0.0071e+12
2.1875e+12 +/-0.0073e+12
+0.9% (XPcal)
-3.1% (XPcal)
Fig.16-19
=> Fig.1-4 in this post
1/H_DARM 4.522e-13 +/- 0.048e-13
4.709e-13 +/- 0.050e-13
4.391e-13 +/- 0.014e-13
4.572e-13 +/- 0.015e-13
N/A
A_BS(1)/A_BS(0) 1.0423 +/- 0.0048 same as left +0.5% (Free-swing) Fig.20-21
A_BS(1) 6.441e-11 +/- 0.050e-11 6.254e-11 +/- 0.057e-11 +1.3% (Free-swing)
H_MICH 4.628e+08 +/- 0.036e+08 4.767e+08 +/- 0.044e+08 +9.8% (Free-swing) Fig.22-25
1/H_MICH 2.161e-09 +/- 0.017e-09 2.098e-09 +/- 0.019e-09 N/A
A_PRM(1)/A_BS(1) 27.21 +/- 0.12 same as left +1.0% Fig.26-27
M2P FF gain 0.03675 +/- 0.00016 same as left N/A
A_PRM 8.764e-10 +/- 0.078e-10 8.509e-10 +/- 0.086e-10 +2.3% (Free-swing)
H_PRCL 2.057e+09 +/- 0.018e+09 2.118e+09 +/- 0.021e+09 +4.3% (Free-swing) Fig.28-31
1/H_PRCL 4.862e-10 +/- 0.043e-10 4.720e-10 +/- 0.048e-10 N/A
A_ITMX(3)/A_BS(1) 0.05369 +/- 0.00049 same as left N/A
M2D FF gain 18.6 +/- 3.2 same as left N/A
YPCAL/XPCAL N/A 0.98474 +/- 0.00080 -0.2% Fig.5-6 in this post

Images attached to this comment
Shingo Hido - 14:07 Tuesday 01 July 2025 (34424) Print this report

I estimated and updated the parameters related DARM (Fig.1).
There is no strange variations (Fig2).

 H_etmxtm = 3.839154366e-14 at 10Hz (0.23% from previous measurements)
 H_etmxim  = 1.535938431e-14 at 10Hz (-1.9% from previous measurements)
 Optical_gain = 2.129473235e12 (0.07% from previous measurements)
 Cavity_pole = 18.127587391 Hz (-0.76% from previous measurements)

Previous parameters are listed in klog#34393.

I investigated the variation in optical gain (specifically, the gain of the sensing function) between 0602 full calibration (klog#34033) and 0630 full calibration (Fig.3).
A decreasing trend was observed from the measurement date (June 2) toward June 20, followed by an increase thereafter.
Since a continuous change can be observed on this time scale (although further data accumulation is needed),
there is a possibility that calibration lines could not only account for statistical fluctuations, but also help suppress systematic errors by correcting for actual variations.

Images attached to this comment
takahiro.yamamoto - 15:27 Tuesday 01 July 2025 (34425) Print this report

MICH and PRCL calibrations were updated by using the values in klog#34423 as follows.
- 1/H_MICH [K1:CAL-CS_SUM_MICH_ERR_FM1]
- A_BS(1) [K1:CAL-CS_SUM_MICH_BS_TM_FM1]
- 1/H_PRCL [K1:CAL-CS_SUM_PRCL_ERR_FM1]
- A_PRM [K1:CAL-CS_SUM_PRCL_PRM_TM_FM1]
Actual changes were shown in Fig.1.

After then line tracking parameters were also updated based on the DARM calibration parameters in klog#34424.
Updated EPICS records can be found in JGW-L2314962.
They were accepted on {observation,down,safe}.snap in klog#34426.

Images attached to this comment
takahiro.yamamoto - 15:41 Tuesday 01 July 2025 (34428) Print this report

Finally, I raised CFC_LATCH bit and moved from CALIB_NOT_READY to READY.

Differences in foton filters shown in Fig.1 are related to klog#34424 (K1:CAL-CS_DARM_*) and klog#34425 (K1:CAL-CS_SUM_{MICH,PRCL}_*).
Differences in SDF tables shown in Fig.2 are related to klog#34411 (K1:CAL-MEAS_*) and klog#34426 (K1:CAL-CS_TDEP_*).
There is no change in guardian code and model files as shown in Fig.3 and Fig.4, respectively.

Images attached to this comment