[kTanaka, Ushiba]
We noticed that the phasing of REFL PDA2 RF51 and RF135 did not seem to be done with PRMI.
So, we checked them.
Figure 1 shows the spectra of REFL_PDA2_RF_51_{I,Q} signals with 94.125 Hz MICH excitation..
The references show the spectra before the phase rotation (102.22 degrees) and the current values show the spectra after the phase rotation (26.82 degrees).
Figure 2 shows the spectra of the REFL_PDA2_RF_135_{I,Q}signals with 94.125 Hz PRCL excitation.
REF0 and REF1 show the spectra before the phase rotation (86.92 degrees) and REF6 and REF7 show the spectra after the phase rotation (129.92 degrees).
So, PD phasing seemed to be not good for both PRCL and MICH.
Since this is the first time we have performed phasing after O4a with PRMI, we are not so sure when the mis-phasing occurred.
After phasing, we tried PRFPMI lock but there are still PRMI glitches are existing when the arms are flashing.
So, further investigation is necessary.
Note:
During phasing, we noticed that MICH has a large offset.
Figure 3 shows the MICH error signals and POP90 signals after RF51 phase rotation.
As you can see, POP90 signals became obviously large when the MICH error signal was changed from 0 to -20 (roughly in units of nm).
We are not so sure what is the source of this offset but if it comes from RF AM, it could be a critical because REFL PD power was drastically changed when arms are flashing for LOCK acquisition and it would cause large offset change of MICH.
We kept new phase values even though PRFPMI could not be locked.
So, if you find some problems, do not hesitate to revert them.
In addition, we also keep K1:LSC-MICH1_OFFSET 20 to get good POP90 with current RF51 phase.
So, if you revert the phase values, please also revert the MICH offset values.
Also, we didn't measure the PRCL and MICH OLTF after the phase rotation, so it is necessary to measure them.