I performed OMMT2 health check.
Figure 1-7 show the TFs from each DoF.
Compared with the TFs in 2023 (klog25967), they look healthy.
Note:
Same as OMMT1 (klog30052), all actuator signs are flipped from the nominal ones.
So, we need to flip the sign of COILOUTF gains.