ryutaro.takahashi - 12:06 Wednesday 20 March 2024 (28909) Print this report
Comment to Installation of side stoppers for MCe (28901)

(Log on 19th)

[Takahashi, Hirata, Washimi, Ikeda]

We recovered the fallen magnet.

First, we aligned the mirror. The yaw pico-motor was set around the mid-point. Then the Oplev spot was far from the QPD. We twisted the mirror in yaw pushing the mirror by the stoppers to the HR direction. After the alignment, we fixed the mirror with the stoppers. We attached the new magnet to the mirror with the gluing jig.