Fabian, Fujii,
= PRM =
-- Based on yesterday's measurement (klog#8306), it was suspected that IM was touching to IMR.
-- Then we found the following three stuffs:
1. IM-H1 flag was touching to the baody horizontally.
2. Some cables out of 4 which is from IM (cables for TM-OSEMs) were touching to IRM gently.
3. Vertical OSEM flags were touching to the OSEM-body, (the flags were shifted horizontally.)
== Tuning work ==
-- For 1, the panel of IM-H1 was shifted so that the flag came to the center of the OSEM cavity.
-- Then we measured a trasfer function (TF) from IML to IML but the TF was still wierd.
-- For 2, we removed one unsued cabe clamp on IMR(see fig1 and fig2.) in ordet to avoid touching
-- Then we measured a trasfer function (TF) from IML to IML but the TF was still wierd.
-- For 3, We have changed the vertical OSEMs-plate horizontally. This was done for V1/V2/V3.
-- Then we measured a trasfer function (TF) from IML to IML and the TF was healthy this time!
== Next ==
-- Tune the vertical OSEMs vertically, for V1 and V2.
-- Measure all the measureable transfer functions for the healthy check.
== Note ==
-- The photos are stored in:
https://photos.app.goo.gl/PUex34gNQ2ANobqy7