MIF (ITF Control)
yoichi.aso - 14:27 Tuesday 26 November 2019 (11756) Print this report
AS RF PD DC calibration

Aso, Nakano

We measured the DC calibration factors of the AS RF PDs with the ITMX single bounce configuration.

We used the Thorlabs's thermal power meter (S310C).

AS PDA1:  4.4mW -> 1940cnts => 2.3e-3 mW/cnts

AS PDA2: 2.0mW -> 1360cnts => 1.5e-3 mW/cnts

Comments to this report:
yuta.michimura - 17:41 Tuesday 26 November 2019 (11762) Print this report

[Aso, Nakano, Michimura]

We measured the optical throughput from BS output to the DARM photodiode (AS_PDA1).
Using independently measured PRM, ITMX and BS reflectivity/transmission, the measured value was about 4.6 %.
We have unknown 60% optical loss.

Details:
- Measured the power at AS PDA1 and PDA2 with ITMX single bounce (see attached).

- From K1:LAS-POW_IMC_DC_OUTPUT, the IMC output power during the measurement was 3.7 W, which gives 0.38 W at BS, assuming PRM transmission is 10.35% (JGW-L1605744) and there's no other large loss. This also gives 95 mW at BS output in ITMX single bounce.

- The measured power at AS PDA1 and PDA2 was 4.4 mW and 2.0 mW, respectively, as reported in klog #11756.

- This means that the optical throughput from BS output to AS PDA1 is 4.4 mW/95 mW = 4.6%.

- Since we have T=30% SRM, we don't currently have OSTM installed, and we have two beamsplitters in the AS table before PDA1, the expected optical throughput is 0.3*0.5*0.5 = 7.5%.

- This means that we have unknown 4.6/7.5 = 60% loss, which sounds kind of OK since there's OFI, viewports etc.

Next:
- Update the shot noise of the noise budget

Images attached to this comment