Reports 1-1 of 1 Clear search Modify search
VIS (IY)
yoichi.aso - 22:49 Thursday 01 August 2019 (9765) Print this report
ITMY Type-A released and roughly aligned

Fujii, Sato, Aso (VIS)
Ushiba, Yamada, Tomaru (CRY)

Bottom Line

  • The Type-A system for ITMY is now released.
  • The alignment is recovered using the OpLev as a reference.
  • With this state, we observed Michelson flinges. So the alignment should be roughly OK.

TO DO

  • Adjust the EQ stop distance for the cryo-payload (Ushiba, Tomaru, Yamada)
  • Clean inside the chamber around the cryo-payload (Ushiba, Tomaru, Yamada)
  • Health check of the suspension with TF measurements. (Lucia, Aso)

Details

Today, CRY people cleaned yet another moving mass then released the cryo-payload.

Then the VIS team tried to release the tower part, finding that some of the IP LVDT signals are freaky.

It was caused by loose connection of in-vac cables. We ended up crimping the D-Sub pins again.

Then we tried to center the IP with the motorized sliders. L and T can be zeroed, but Y was difficult to zero, because it will require one of the sliders to be very close to the edge of the moving range. We decided to forget about putting Y to zero, because we can adjust Yaw with the stepper motor. IP legs tend to have their own confortable equilibrium points. Even if that position creates a large Yaw signal in the LVDTs, we don't care, as long as LVDTs are not close to saturation. We only need to put L and T zero.

The current set points for the IP control are: L = 100, T=200, Y=-710
#L=100 and T=200 comes from when Fujii-kun searched for an IP position which makes the BF Yaw transfer function better

After releasing the IP, we tried to recover the mirror alignment with OpLev.

Along the way, we found that the weight of the cryo-payload changed from the addition of the wire clamp, causing the change of the GAS DC positions. CRY people adjusted the weight with ballast masses also adjusting pitch. 

Finally, we adjusted Yaw with the F0 Yaw stepper motor and fine adjusted Pitch with the moving masses.

After the OpLev is back to the reference point for the alignment, we removed the aluminum foil to block the beam into the IYC chamber.

Voila !! The Michelson flinge was there !

We confirmed that the alignment was roughly OK. We called it a day.

Comments to this report:
yoshinori.fujii - 8:54 Friday 02 August 2019 (9766) Print this report

Supplementary note

  • > Then the VIS team tried to release the tower part, finding that some of the IP LVDT signals are freaky.
    > It was caused by loose connection of in-vac cables. We ended up crimping the D-Sub pins again.
    • We did this work for all the IP-LVDTs (H1/H2/H3), and confirmed that all of them were electrically connected even when we touched the connectors by hand.This isssue went away for ITMY suspension.
  • > We decided to forget about putting Y to zero, because we can adjust Yaw with the stepper motor.
    • Just for making sure; because we can adjust "TM-Yaw" with the stepper motor "on F0-GAS for Yaw".
  • > The current set points for the IP control are: L = 100, T=200, Y=-710
    > #L=100 and T=200 comes from when Fujii-kun searched for an IP position which makes the BF Yaw transfer function better
    • For L and T, this set point was used since the transfer function of BF-Yaw was optimized at that point so far (klog#9172), however, there is not much reason to stick the IP at this point if the transfer functions are healthy. If healthy TFs are measured at ex. (L, T) = (0, 0), we can change the IP position to the point.
      • The point was totally depended on the BF-Yaw TF, thus we should want to check it with changing IP position to find the sweet point.
    • For Y, this position was decided so that the IP-motor-H2 had some margin. However, after the work, I found that the current margin might not be enough for the pumping-down compensation business, according to a measurement in klog#9732. (see the Figure 1 to 3, more beautiful ones are to be posted.) This confirmation was done by eye at the site during the buisiness and I seemed to find a gap between the limitter and the motor, however, it might not be enough.
      • The current motor set point after the IP centering work was (H1, H2, H3) = (43.0mm, 36.0mm, 43.0mm). The definition is in Figure 4.
      • Thus the Yaw position might be necessary:
        • to change the set point in Yaw so that the IP-motor-H2 has larger mergin.
        • or, to change the motor limmiter position
      • FYI, until quite recently the IP set point was at (L, T, Y) = (100, 200, 0), but it looked that all the IP-slider-motors were at the edge of the motor-range. NOT setting Yaw to zero should save us.
      •  we also confirmed that the accelerometers, 3 of them, became operational successfully, even though we did not check the noise levels in spectra. Still it looked noisy by eye.
  • > After releasing the IP, we tried to recover the mirror alignment with OpLev.
    • This was done with F0(Y) motor since the picomotor on the BF for Yaw adjustment was not available.
    • The oplev was back when IP was at (100, 200, -710) and BF-Yaw was at 11350.
    • We decide that we do not tune the BF-LVDT position horizontally at that time since the position was not so close to the secondary coil holders. If the IP position is tuned again, we might have to takecare of this though.
  • > Along the way, we found that the weight of the cryo-payload changed from the addition of the wire clamp, causing the change of the GAS DC positions. CRY people adjusted the weight with ballast masses also adjusting pitch. 
    • During this work, we also found that the sensor polarity of the F1- and BF- GAS were NOT collect.. This is to be fixed maybe after the ITMY pumping is finished(?) in order to avoid confusions.
    • We adjusted some amount of masses so that the BF-GAS filter oscillates. A compensation with fishing rods is necessary.
    •  
yoshinori.fujii - 9:07 Friday 02 August 2019 (9767) Print this report

Attachments.

Only the coeeficient is useful.

Note:

For F0Y, a negative value makes positive yaw direction. The red/blue curve is the results before/after the PEEK part was installed.

Images attached to this comment
yoshinori.fujii - 9:13 Friday 02 August 2019 (9768) Print this report

Summary of the next steps:

  • take care of the IP-H2 motor
  • GAS position tuning with the fishing rods.
  • conduct the mechanical transfer function measurement and confirm if the suspension is healthy indeed [1]
    • for the template files, see klog#9734.
    • /users/VISsvn/TypeAtower/ITMY/Measurements/TF/template/
    • /users/VISsvn/TypeAtower/ITMY/Measurements/TF/reference/

[1] After the measurement, do the followings:

  • Compare the results to the references, i.e., to the simulated ones and previous measurements:
    • Check if a resonant peak and Q-factor at its frequency are as we designed.
    • Check if the dc-gain is comparable with the simulated one, previous or the other suspensions'
    •  
    • NOTE:
    • Especially IP-L, T and BF-Yaw d.o.fs. is necessary to be chacled deeply.
    • Small band-width like 3mHz will be necessary, and smaller excitation amplitude is recommended in that measurement in order NOT to damage the suspensions.
[1] After the measurement, do the followings:
  - Compare the results to the references, i.e., to the simulated ones and previous measurements:
  - Check if a resonant peak and Q-factor at its frequency are as we designed.
  - Check if the dc-gain is comparable with the simulated one, previous or the other suspensions'
    - Especially IP-L, T and BF-Yaw d.o.fs.
    - Small band-width like 3mHz will be necessary, and smaller excitation amplitude is recommended in that measurement in order NOT to damage the suspensions. 
Search Help
×

Warning

×