The sideband frequency of the IMC was 51.72 MHz, not 52.72 MHz. So, the caribration was wrong.
The actual measured FSR is 5.62 +/- 0.02 MHz, and the roundtrip length is 53.4 +/- 0.2 m, which is consistent with the design.
Sorry we made a mistake!