[Hirose, Chen]
We increased the gain of RPD1 (TR IR PD) on the TMSX optical table from 0 dB to 40 dB.
To verify the gain change, we illuminated the PD with a work light and compared the signal levels before and after the adjustment.
We confirmed that the signal amplitude increased by a factor of approximately 100, consistent with the expected gain increase of 40 dB.