[Fujimoto, Komori]
We started investigating whether shifting the RF sideband frequency could improve the MICH signal.
As a first trial, we decreased the seed frequency by 300 Hz. We then adjusted the demodulation phase of RF51 to -170 degree so as to minimize the SRCL contribution in the Q phase, and measured the MICH response.
The attached figure shows the result of the MICH excitation, with the blue trace representing the Q-phase spectrum.
The peak-to-floor ratio is approximately 10, which is higher than the value of approximately 4 obtained in the previous measurement (klog:37832).
This result suggests that the MICH signal may be recovered by adjusting the SRCL.
Tomorrow, we will perform the decoupling and attempt to lock the DRMI using the 3f signals with this sideband frequency.