[Ushiba, Hirose, Aritomi, Fujimoto, Kataoka, Sugioka]
Abstract:
To construct input matrix for DRMI3F and decouple each signals, we measured coupling levels.
We then assessed the 3f signals and attempted the 1f-to-3f handover.
The handover failed, and it is likely that the MICH signal is still not sufficiently decoupled.
Details:
We measured the MICH-, PRCL-, and SRCL-to-3f PD signal transfer functions (TFs) while DRMI 1f was locked.
We injected a 150.125 Hz excitation into each DoF 1f control loop and measured the TF from the 1f IN1 signal to the 3f PD signal.
Since the IN1 signal is calibrated in nanometers, the gain is calibrated in dBcnt/nm.
The measurements were performed with the REFL51 demodulation phase set to 145(Figure.1) and 130(Figure.2)(as detailed in klog37830).
The left/middle/right pads shows the measurements under PRCL/SRCL/MICH excitation, respectively.
In both demodulation phase cases, the MICH-to-REFL51Q gain was remarkably small even compared to MICH-to-REFL51I.
This might imply that MICH signal is either in phase or in antiphase with PRCL and SRCL.
Based on these measurements, we constructed the 3f control signal and evaluated it. This part was performed exclusively with a demodulation phase of 145°.
We constructed a 3*3 matrix with nine each measured TF gain, and used the inverse of this matrix for Input Matrix.
We utilized the CARM, DARM, and XARM IN1 channels—which are not currently used for control—as the 3f MICH,PRCL, and SRCL signals.
Then we injected 150.125Hz excitation to each DoF 1f contorl loop again, and measured TF from 1f IN1signal to 3f IN1 signal(Figure.3).
Figure.3 shows that 3f PRCL and SRCL signal is well decoupled.
Although the 3f MICH signal exhibits some coupling from PRCL and SRCL, its coherence is low.
We also measured the OLTF of these 3f signals while DRMI1f locked(Figure.4 for MICH, Figure.5 for PRCL, Figure.6 for SRCL) which generally matched to those of 1f.
Subsequently, we attempted the 1f-to-3f handover.
First, we tried the MICH handover, but it was unsuccessful.
Next, we attempted to hand over PRCL and SRCL first, followed by MICH control.
The PRCL handover—which had already been achieved yesterday—was successful, but PRCL 3f was unstable, and the SRCL handover attempt was skipped.
Plans:
There are several potential approaches for DRMI 3f locking:
-
The surprisingly weak MICH signal in REFL51Q is unusual; therefore, further investigation was conducted, which will be posted by Hirose-san later.
-
Taking the SRCL signal from another port and using REFL51I for MICH should also be feasible.
-
There is a possibility that imperfect phasing and decoupling in 1f compromised the matrix adjustment. Consequently, further fine-tuning of the 1f control could serve as a solution.