Hiroki Fujimoto - 2:28 Saturday 14 November 2026 (37842) Print this report
Investigation of MICH Signal in REFL51-I and REFL135-Q

[Sugioka, Hirose, Komori (online), Ushiba, Fujimoto]

Abstract

We investigated whether the MICH signals observed in REFL51-I and REFL135-Q were actual MICH signals or originated from coupling due to imperfect decoupling of the MICH actuator.
As a result, we found that the MICH signal observed in REFL51-I was likely due to coupling.
On the other hand, the estimated coupling contributions in REFL135-Q were more than one order of magnitude smaller than the observed signal, suggesting that REFL135-Q was likely sensing the actual MICH signal.
In addition, the signal-to-noise ratio relative to the noise floor was about six times better in REFL135-Q than in REFL51-Q, suggesting that REFL135-Q could be used for MICH control.
However, as reported in ⁠klog 37835, REFL135-Q also has coupling from PRCL and SRCL. In particular, the PRCL coupling is larger than the MICH response. Therefore, decoupling by subtracting the coupled signals is essential.

Details

Previous investigations of signal coupling in the DRMI suggested that MICH signals might appear in REFL51-I and REFL135-Q.
However, when measuring the MICH response of each sensor, we used the MICH actuator, which combines the BS, PRM, and SRM actuators. If the MICH actuator was not sufficiently decoupled, it could also excite PRCL and SRCL. Therefore, the observed signals might have originated from coupling through PRCL and SRCL rather than from MICH itself.

To investigate this possibility, we estimated the excitation amplitudes of PRCL and SRCL when injecting a signal into the MICH actuator, using the 1f sensors.
We then multiplied these amplitudes by the transfer functions from the 1f sensors to the 3f sensors to estimate the peak amplitudes appearing in REFL51-I and REFL135-Q through PRCL and SRCL coupling.
By comparing these estimated amplitudes with the measured amplitudes, we evaluated whether the observed signals originated from MICH itself or from coupling.

Figs. 1, 2, and 3 show the spectra of the signals when excitations were injected into MICH, PRCL, and SRCL, respectively, while the DRMI was locked with the 1f sensors.
We used the peak amplitudes and transfer functions obtained from these measurements to estimate the coupling contributions.

MICH to REFL51-I

The measured peak amplitude in REFL51-I when exciting MICH and the estimated peak amplitudes due to coupling through PRCL and SRCL are as follows:

  • Measured peak amplitude in REFL51-I: 0.074
  • Estimated peak amplitude through PRCL: 0.022
  • Estimated peak amplitude through SRCL: 0.031

Since the estimated coupling contributions are of the same order of magnitude as the measured peak amplitude, the signal observed in REFL51-I when exciting MICH is likely dominated by coupling.
However, if the relative phase between the PRCL and SRCL coupling contributions is close to 180 degrees, these contributions could cancel each other out.
Therefore, we cannot rule out the possibility that most of the observed signal is actually a genuine MICH signal.

(The following values were used to calculate the estimated peak amplitudes:

  • Peak amplitude in PRCL_IN1: 0.015
  • Transfer function from PRCL_IN1 to REFL51-I when exciting PRCL: 3.16 dB
  • Peak amplitude in SRCL_IN1: 0.11
  • Transfer function from SRCL_IN1 to REFL51-I when exciting SRCL: -11 dB)

MICH to REFL135-Q

The measured peak amplitude in REFL135-Q when exciting MICH and the estimated peak amplitudes due to coupling through PRCL and SRCL are as follows:

  • Measured peak amplitude in REFL135-Q: 0.023
  • Estimated peak amplitude through PRCL: 0.0017
  • Estimated peak amplitude through SRCL: 0.0019

Since the estimated coupling contributions are more than one order of magnitude smaller than the measured peak amplitude, the observed signal is likely a genuine MICH signal rather than a signal originating from coupling.
Therefore, REFL135-Q could potentially be used for MICH control.
However, as reported in ⁠klog 37835, REFL135-Q also has coupling from PRCL and SRCL. In particular, the PRCL coupling is larger than the MICH response. Therefore, decoupling by subtracting the coupled signals is essential.

(The following values were used to calculate the estimated peak amplitudes:

  • Peak amplitude in PRCL_IN1: 0.015
  • Transfer function from PRCL_IN1 to REFL135-Q when exciting PRCL: -19.1 dB
  • Peak amplitude in SRCL_IN1: 0.11
  • Transfer function from SRCL_IN1 to REFL135-Q when exciting SRCL: -35.3 dB)
Images attached to this report
Comments to this report:
Tatsuya Sugioka - 23:21 Saturday 14 November 2026 (37843) Print this report

We further investigated the MICH signal observed in REFL51.

Estimation of Coupling Contributions Including Phase Information

First, we estimated the coupling contributions from PRCL and SRCL, taking into account the phases of their TFs.
For REFL51I, the phases of the coupling TFs through the PRCL and SRCL paths are 127° and −37°, respectively, which are nearly opposite in phase.
As a result, the two coupling contributions largely cancel each other, and the estimated total coupling gain for MICH-to-REFL51I is −31.2 dB, whereas the measured gain is −15.4 dB.
This suggests that the observed signal in REFL51I is dominated by the genuine MICH response rather than the actuator coupling contributions via PRCL and SRCL.

Dependence on the Demodulation Phase

Next, we estimated the coupling contributions at various REFL51 demodulation phases.
As described in klog37830, the PRCL and SRCL signals in REFL51Q are minimized at demodulation phases of 130° and 145°, respectively.
Using this information, we calculated the PRCL/SRCL coupling contributions to REFL51I/Q as functions of the demodulation phase.
We also measured the MICH-to-REFL51I/Q TFs by exciting MICH(Figure.1) and compared the measured gains with the estimated coupling contributions (Figure.2).
The direct coupling contributions from PRCL and SRCL are also shown.
The results indicate that the estimated actuator coupling contributions are negligible compared with the measured MICH-to-REFL51 signals, except where the measured gains fall below approximately −40 dB.
However, the response to MICH is much weaker than that to PRCL and is comparable in magnitude to the SRCL response.

According to the optical design, the MICH and SRCL signals are expected to be well separated in the I/Q plane.
However, the measured responses are nearly aligned, with a phase difference of only approximately 5°.
The reason for this discrepancy is not yet understood.

Note: We cannot rule out the possibility that the observed "MICH signal" contains additional coupling contributions that have not been considered in our estimation.
If such contributions are significant, the genuine MICH response could be much smaller than the measured value.

Images attached to this comment