[Sugioka, Hirose, Komori (online), Ushiba, Fujimoto]
Abstract
We investigated whether the MICH signals observed in REFL51-I and REFL135-Q were actual MICH signals or originated from coupling due to imperfect decoupling of the MICH actuator.
As a result, we found that the MICH signal observed in REFL51-I was likely due to coupling.
On the other hand, the estimated coupling contributions in REFL135-Q were more than one order of magnitude smaller than the observed signal, suggesting that REFL135-Q was likely sensing the actual MICH signal.
In addition, the signal-to-noise ratio relative to the noise floor was about six times better in REFL135-Q than in REFL51-Q, suggesting that REFL135-Q could be used for MICH control.
However, as reported in klog 37835, REFL135-Q also has coupling from PRCL and SRCL. In particular, the PRCL coupling is larger than the MICH response. Therefore, decoupling by subtracting the coupled signals is essential.
Details
Previous investigations of signal coupling in the DRMI suggested that MICH signals might appear in REFL51-I and REFL135-Q.
However, when measuring the MICH response of each sensor, we used the MICH actuator, which combines the BS, PRM, and SRM actuators. If the MICH actuator was not sufficiently decoupled, it could also excite PRCL and SRCL. Therefore, the observed signals might have originated from coupling through PRCL and SRCL rather than from MICH itself.
To investigate this possibility, we estimated the excitation amplitudes of PRCL and SRCL when injecting a signal into the MICH actuator, using the 1f sensors.
We then multiplied these amplitudes by the transfer functions from the 1f sensors to the 3f sensors to estimate the peak amplitudes appearing in REFL51-I and REFL135-Q through PRCL and SRCL coupling.
By comparing these estimated amplitudes with the measured amplitudes, we evaluated whether the observed signals originated from MICH itself or from coupling.
Figs. 1, 2, and 3 show the spectra of the signals when excitations were injected into MICH, PRCL, and SRCL, respectively, while the DRMI was locked with the 1f sensors.
We used the peak amplitudes and transfer functions obtained from these measurements to estimate the coupling contributions.
MICH to REFL51-I
The measured peak amplitude in REFL51-I when exciting MICH and the estimated peak amplitudes due to coupling through PRCL and SRCL are as follows:
- Measured peak amplitude in REFL51-I: 0.074
- Estimated peak amplitude through PRCL: 0.022
- Estimated peak amplitude through SRCL: 0.031
Since the estimated coupling contributions are of the same order of magnitude as the measured peak amplitude, the signal observed in REFL51-I when exciting MICH is likely dominated by coupling.
However, if the relative phase between the PRCL and SRCL coupling contributions is close to 180 degrees, these contributions could cancel each other out.
Therefore, we cannot rule out the possibility that most of the observed signal is actually a genuine MICH signal.
(The following values were used to calculate the estimated peak amplitudes:
- Peak amplitude in PRCL_IN1: 0.015
- Transfer function from PRCL_IN1 to REFL51-I when exciting PRCL: 3.16 dB
- Peak amplitude in SRCL_IN1: 0.11
- Transfer function from SRCL_IN1 to REFL51-I when exciting SRCL: -11 dB)
MICH to REFL135-Q
The measured peak amplitude in REFL135-Q when exciting MICH and the estimated peak amplitudes due to coupling through PRCL and SRCL are as follows:
- Measured peak amplitude in REFL135-Q: 0.023
- Estimated peak amplitude through PRCL: 0.0017
- Estimated peak amplitude through SRCL: 0.0019
Since the estimated coupling contributions are more than one order of magnitude smaller than the measured peak amplitude, the observed signal is likely a genuine MICH signal rather than a signal originating from coupling.
Therefore, REFL135-Q could potentially be used for MICH control.
However, as reported in klog 37835, REFL135-Q also has coupling from PRCL and SRCL. In particular, the PRCL coupling is larger than the MICH response. Therefore, decoupling by subtracting the coupled signals is essential.
(The following values were used to calculate the estimated peak amplitudes:
- Peak amplitude in PRCL_IN1: 0.015
- Transfer function from PRCL_IN1 to REFL135-Q when exciting PRCL: -19.1 dB
- Peak amplitude in SRCL_IN1: 0.11
- Transfer function from SRCL_IN1 to REFL135-Q when exciting SRCL: -35.3 dB)