ISC (IMC)
Seidai Kawakami - 21:13 Wednesday 21 October 2026 (37713) Print this report
Results of MCE adjustment

[Takahashi, Aso, Komori, Washimi, Miyamoto, Kawakami]

Continued from klog37708

Summary

We adjusted the position of MCE using pushers and side clamps.
Also, we checked the beam path of the oplev after the adjustment, and we carried out oplev centering.
MCE was successfully moved toward MCF by an average of 10.225±0.005 mm, close to the target displacement of 10.26 mm.

Details

・First, we used side clamps to prevent MCE from shifting laterally and used pushers to move it toward MCF. The target displacement was 10.26 mm.
   Figures 1 and 2 show the interior of the MCE chamber.
・We measured the displacement by measuring the distances between the side clamp on the measurement side and the MCE base at the left and right side.
   The measurement results, as viewed from the MCF side, are listed below. The distances were measured using a caliper with a resolution of 10 µm.
   

Left-side distance[mm] Right-side distance[mm]    Remarks   
3.72 6.52 Initial position
3.71 6.51 Installed the pusher
4.58 7.45 Turn the pusher 3 full turn
7.63 10.37 Turn the pusher 10 full turns
10.69 13.41
10.78 13.52  
13.82 16.56 Turn the pusher 10 full turns
13.82 16.64 Remeasurement
13.82 16.56
13.86 16.60 Turn the pusher by 45 degrees
13.92 16.66
13.97 16.75
13.97 16.68 Removed the pusher
13.97 16.65  Remeasurement 
13.97 16.63
13.97 16.80 Installed the pusher
13.98 16.73 Remeasurement
13.92 16.71
13.96 16.66 Secured one point
13.95 16.70 Removed the pusher
13.98 16.66 Secured one point 
14.00 16.76 removed the clamp and turn pusher by 45 degrees
13.99 16.67 Secured one point
14.01 16.68 Secured two point
14.02 16.63 Secured three point
14.02 16.67 Secured four point


・After securing the MCE base in its final position, the displacements were 10.30 mm on the left side and 10.15 mm on the right side, giving an average MCE displacement of 10.225 mm.
・The readings on the right side varied irregularly between measurements, but the cause is unknown.
 

 

Images attached to this report
Comments to this report:
Seidai Kawakami - 9:35 Thursday 22 October 2026 (37718) Print this report

There are two corrections to this report.

  1. The uncertainty in the displacement was initially reported as 0.005 mm. However, the readings on the right side showed a scatter of approximately 0.1 mm, so an uncertainty of 0.1 mm was assigned to the right-side measurement. Since the final displacement was calculated by averaging the left- and right-side displacements, the resulting uncertainty was 0.05 mm.

  2. In the measurement table, the Remarks entry for the row with a left-side distance of 13.95 mm and a right-side distance of 16.70 mm should read “Removed the clamp” instead of “Removed the pusher.”