takahiro.yamamoto - 10:07 Thursday 15 October 2026 (37672) Print this report
Hang-up of models on K1MCF0
Real-time models on K1MCF0 hung-up at 01:01:28 JST.
Affected models are k1iopmcf0, k1pemmcf0, and k1pemmanage.

Because K1IMC0 is alive, it's not related to the problem on the power supply unit reported as klog#37524.
Though I'm not sure the large speaker was enabled as the recovery work from the unexpected blackout on Sep. 21st.
it may be a same issue as klog#25343, klog#29433, and klog#33565.
Comments to this report:
takahiro.yamamoto - 18:13 Thursday 15 October 2026 (37677) Print this report
Upon visual inspection of the MCF0 rack, the following findings were confirmed:
  • The speaker amplifier is OFF.
  • The 18 V and 24 V voltages measured at the power strips are within the normal ranges.
  • All circuits appear to be operating normally, and no abnormalities were observed in the indicators.
  • No abnormalities were observed in the indicators inside the I/O chassis.
  • The 3 V, 5 V, and 12 V voltages measured on the boards are all within the normal ranges.
  • The Timing Slave indicator indicates that synchronization is in progress.

Additionally, since the Real-Time Front-End (RTFE) is accessible via login, it was confirmed that the kernel is still responsive. Furthermore, the lspci command showed that the PCIe devices in the I/O chassis are not detected. This suggests that the issue is not simply an ADC timeout caused by a timing glitch. Therefore, a disconnection or failure in the PCIe link between the I/O chassis and the RTFE is suspected.

At this stage, we have not yet determined whether the issue is related to the 24 V power supply, including the power supplied to the boards inside the I/O chassis, or to the PCIe connection itself. If the issue is power-related, it may be necessary to shut down the IMC0 rack, which shares the same power supply unit (PSU) as the I/O chassis.

For this reason, we have decided to defer recovery operations until the scheduled maintenance window on Monday.