VAC (OMMT)takashi.uchiyama - 12:37 Wednesday 16 September 2026 (37490)
Print this reportComment to Opened vcauum chamber OMMT (37429)2026/09/16
mTakahashi, Sawada, Uchiyama
We closed two OMMT flanges to the proper torque: 10 ->20 -> 40 -> 40 N*m. After that, we started pumping down.
9:58, Start pumping down using a dry pump at OMMT. We opened 1/2 turn of the angle valve between the dry pump and OMMT. 10:57, Reached 0.01 atm. We opened the angle valve fully. 11:56, Vacuum pressure reached 20 Pa. We started changing the dry pump to TMP at SRM. 12:30, TMP reached normal operation.