Abstract
For the DRMI LSC, I reduced the coupling from PRCL to the MICH signal and diagonalized the MICH actuator.
As a result, the strange shape of the 3f OLTFs reported previously was resolved, and cleaner OLTFs were obtained.
The lock duration was also improved, and the 3f lock was maintained for more than 11 hours (still locked now).
These settings have not yet been implemented in the guardian. I plan to implement them after they are reviewed by other commissioners.
Details
In the previous DRMI 3f lock, as reported in klog #37346, the shapes of the PRCL and MICH OLTFs after the transition to 3f were strange.
This suggests that there was coupling between the degrees of freedom.
Therefore, for the DRMI 3f lock, I performed sensor diagonalization and actuator diagonalization and also adjusted the gains of the OLTFs.
During this 3f lock, the ADS loops for BS and SRM were turned off.
Removal of PRCL coupling
After locking the DRMI with the 3f signals, I used OSC1 in LSC_OVERVIEW with the following settings and excited PRM in the length direction above the control bandwidth.
- Amplitude: 6000
- Frequency: 230 Hz
- Injection point: PRM
In this state, I checked the peaks in the spectra of the PRCL, MICH, and SRCL error signals.
Then I adjusted the factor applied to REFL135I when constructing the MICH and SRCL signals, in order to remove the peaks.
Fig. 1 shows the peaks after reducing the coupling from PRCL. The factors were changed as follows.
- REFL135I_to_MICH: -1.38 ⇒ -0.75
- REFL135I_to_SRCL: 0.705 ⇒ 0.705 (unchanged)
The coupling to MICH was reduced by about a factor of 7, while the coupling to SRCL could not be reduced further.
In addition, especially for SRCL, the coupling seemed to have some alignment dependence.
In some cases, the coupling slightly increased as shown in Fig. 1, even though the factor was not changed.
Removal of SRCL coupling
In the same way as above, I excited SRM above the control bandwidth and tried to remove the SRCL signal from PRCL and MICH. However, as shown in Fig. 2, the coupling from SRCL was already sufficiently small.
Therefore, I did not perform any further coupling reduction for SRCL.
Diagonalization of the MICH actuator
Previously, only BS was used as the actuator for MICH. However, this also moves PRCL and SRCL.
Therefore, I adjusted the actuation so that only MICH is driven by moving PRM and SRM together with BS.
Specifically, I used OSC1 in LSC_OVERVIEW to drive BS, PRM, and SRM, and adjusted the factors so that the excitation appeared only in the MICH error signal.
The three reference traces in Fig. 3 show the spectra when only BS was excited with an amplitude of 6000 × 15.
The first three traces in the legend show the spectra after the factors were adjusted.
The OUTPUT_MTRX values from OSC1 to BS, PRM, and SRM used in this adjustment were as follows.
- BS: 15
- PRM: -0.6
- SRM: 5.1
With this adjustment, the coupling to SRCL was reduced by about a factor of 5, and the coupling to PRCL was reduced by about a factor of 50.
Using the ratios obtained here, I changed the MICH actuator settings in the OUTPUT_MTRX of LSC_OVERVIEW as follows.
- MICH_to_PRM: 0 ⇒ -0.04
- MICH_to_SRM: 0 ⇒ 0.34
- MICH_to_BS: 1.0 ⇒ 1.0 (unchanged)
OLTFs
After the sensor and actuator diagonalization described above, I measured the OLTFs of MICH, PRCL, and SRCL, and adjusted the overall gains.
Figs. 4, 5, and 6 show the OLTFs of MICH, PRCL, and SRCL after the gain adjustment, respectively.
The overall gains were changed as follows.
- MICH1 gain: 2.1 ⇒ 7.5
- PRCL1 gain: 2.0 ⇒ 2.0 (unchanged)
- SRCL1 gain: 1.4 ⇒ 2.0
With the previous settings, as reported in klog #37346, the shapes of the 3f MICH and PRCL OLTFs were abnormal.
However, as shown in Figs. 4 and 5, probably thanks to the diagonalization performed this time, I obtained relatively clean OLTFs similar to those of the 1f lock.
Lock duration
Even after the sensor and actuator diagonalization and the corresponding setting changes, the DRMI remained locked with the 3f signals.
As shown in Fig. 7, the lock was maintained for more than 11 hours.
Screenshot of the current setup
Fig. 8 shows a screenshot of the current setup.
Summary and future work
This time, I diagonalized the MICH 3f sensor and the MICH actuator.
As a result, the shapes of the 3f OLTFs became cleaner, and the lock also seemed to be maintained more stably.
These settings have not yet been implemented in the guardian.
If these settings are confirmed to be appropriate, I would like to update the guardian settings accordingly.
In addition, the output of LSC-MICH_OUT_DQ was close to 100000 and seemed to be close to saturation.
Therefore, the roll-off filters or related settings should also be adjusted.
For ASC, the ADS loops for BS and SRM were not engaged this time, so they need to be also optimized.
And since the sensor coupling seemed to depend on the alignment, it may be necessary to do the diagonalization procedure described above again after introducing the ADS loops.
In parallel with working on these items, we would like to try the detuned-arm + DRMI 3f lock again.