A CAL Tcam session was performed to obtain beam position information necessary for Pcal. The parameters have already been updated, and SDF has already been accepted.
During this work, we adjusted Pcal-X beam position by using PCAL_EX2 pico. The related SDF has been accepted.
The RxPD output is backed to usual values form the changed value reported on klog35342.
A CAL Tcam session was performed to obtain beam position information necessary for Pcal. The parameters have already been updated, and SDF is expected to be accepted.
Operator: Dan Chen
Update Time: 2025/10/17 08:05:30
| EPICS Key | Before [mm] | After [mm] | Δ (After - Before) [mm] |
|---|---|---|---|
| K1:CAL-PCAL_EX_TCAM_PATH1_X | -0.05083 mm | 1.25360 mm | +1.30443 mm |
| K1:CAL-PCAL_EX_TCAM_PATH1_Y | 64.40961 mm | 64.28063 mm | -0.12898 mm |
| K1:CAL-PCAL_EX_TCAM_PATH2_X | -2.28924 mm | -0.26791 mm | +2.02133 mm |
| K1:CAL-PCAL_EX_TCAM_PATH2_Y | -64.92314 mm | -64.21604 mm | +0.70710 mm |
Update Time: 2025/10/17 08:06:16
| EPICS Key | Before [mm] | After [mm] | Δ (After - Before) [mm] |
|---|---|---|---|
| K1:CAL-PCAL_EY_TCAM_PATH1_X | 1.39792 mm | 1.37528 mm | -0.02264 mm |
| K1:CAL-PCAL_EY_TCAM_PATH1_Y | 64.52843 mm | 64.41668 mm | -0.11175 mm |
| K1:CAL-PCAL_EY_TCAM_PATH2_X | -0.42296 mm | -0.52695 mm | -0.10399 mm |
| K1:CAL-PCAL_EY_TCAM_PATH2_Y | -69.94413 mm | -70.02516 mm | -0.08103 mm |