Reports of 27008
VIS (IMM)
takafumi.ushiba - 9:10 Thursday 25 April 2024 (29320) Print this report
Health check of IMMT2

I performd health check of IMMT2 (fig1 - fig7).
All TFs seem fine.

Images attached to this report
VIS (IMM)
takafumi.ushiba - 9:02 Thursday 25 April 2024 (29319) Print this report
Health check of IMMT1

I performd health check of IMMT1 (fig1 - fig7).
All TFs seem fine.

Images attached to this report
PEM (Center)
takaaki.yokozawa - 7:45 Thursday 25 April 2024 (29318) Print this report
Measurement of the vibration on the vactable of the OMC
[YokozaWashimi, Ozaki, Sudo]

We performed the measurement of the vibration spectrum using the 3axis accelerometer.
Totally, 10 position measurement were performed.
The summary will be appeared soon.
Images attached to this report
VIS (SRM)
ryutaro.takahashi - 0:08 Thursday 25 April 2024 (29317) Print this report
Visual inspection of SRM suspension

[Ikeda, Takahashi]

We checked the OSEMs visually. The flap of OSEM#3 and #4 in the IM was rotated more than 40°.

Images attached to this report
VIS (SR3)
ryutaro.takahashi - 23:46 Wednesday 24 April 2024 (29316) Print this report
Visual inspection of SR3 suspension

[Ikeda, Takahashi]

We checked the OSEMs visually. The flap of OSEM#2 in the IM was rotated more than 40°.

Images attached to this report
VIS (SR2)
ryutaro.takahashi - 23:13 Wednesday 24 April 2024 (29315) Print this report
Visual inspection of SR2 suspension

[Ikeda, Takahashi]

We checked the OSEMs visually. The flap of OSEM#1 in the IM was rotated more than 40°.

Images attached to this report
VIS (IMM)
ryutaro.takahashi - 22:19 Wednesday 24 April 2024 (29314) Print this report
Visual inspection of IMMT2 suspension

[Takahashi, Ikeda, Hirata, Ushiba]

We checked the IMMT2 suspension visually. There was a small margin of the Pico-motor range for the pitch direction. There was no margin for yaw CCW direction (<1mm). There were not any rubbing magnets in the IM.

Images attached to this report
VIS (IMM)
ryutaro.takahashi - 21:50 Wednesday 24 April 2024 (29313) Print this report
Visual inspection of IMMT1 suspension

[Takahashi, Ikeda, Hirata, Ushiba]

We checked the IMMT1 suspension visually. There were margins of the pico-motor range much enough to adjust both pitch and yaw motion. There were not any rubbing magnets in the IM. When we took the pictures with a fiber scope touching the EQ stop frame, the TM pitch jumped due to a weak joint in the X-Y stage supporting the frame.

Images attached to this report
MIF (General)
naoatsu.hirata - 20:56 Wednesday 24 April 2024 (29312) Print this report
Comment to Finalization of IFI-IMM-PRM: Day 4; Re-alignment of IMMT1, IMMT2, and PRM (29299)

[Hirata, Ushiba-san, Ikeda-san]

We checked IR beam position around PR2 HR side Mid-size baffle. (How to make the vertical line is same as klog:29282)
It looks that IR beam positon is about 3.5mm +Y side away from the center of aperture.(IR beam center is 68.5mm and aperture center is 72mm on the ruler.)

Images attached to this comment
FCL (Water)
tatsuki.washimi - 18:13 Wednesday 24 April 2024 (29311) Print this report
Comment to Y-arm mine water and pumps (29047)

The readout of the water fluid has recovered.

Images attached to this comment
VIS (OMM)
tatsuki.washimi - 17:54 Wednesday 24 April 2024 (29310) Print this report
Comment to Visual inspection of Stack (29245)

[YokozaWashimi, Tanaka, Ozaki, Sudo]

Today we tried to evaluate the seismic isolation of the OMC stacks, using a 3-axial accelerometer (S2315344) and an impact hammer (G1910656).
This is a quick report.

  1. Locating the ACC on the OMC vac-table (Fig.1, Fig.2)
  2. Tapping each stack (top, middle, bottom of stack1, stack2) with the impact hammer (direction: Fig.3)
  3. Picking up each pulse (-0.2/+0.8 sec) and calibrating its transfer functions from impact [N] to acceleration [m/s2] automatically (example: Fig.4, Fig.5)
  4. Calculating averages of the events (Fig.6, Fig.7)
Images attached to this comment
LAS (High Power Laser R&D)
tomoyuki.uehara - 16:42 Wednesday 24 April 2024 (29309) Print this report
neoLASE laser
[Miyakawa, Uehara]
We checked some laser operations. And we set a mirror for beam profile measument later(fig1).
Images attached to this report
MIF (General)
takafumi.ushiba - 16:42 Wednesday 24 April 2024 (29308) Print this report
Finalization of IFI-IMM-PRM: Day 5; Reconstruction of POP forward beam path

[Hirata, Ushiba]

Abstract:

We reconstructed POP forward beam path.
Though beam is still close to the knob of mirror mount just after periscope, the beam doesn't seem clipped.
Now, IR beam hits almost center of the both QPDs when the alignment to PR2 is good.

Detail:

1. Check of beam position at high power beam dump:

First, we requested aLIGNED state for IMMT1, IMMT2, and PRM.
Then, to confirm the good alignment has been kept from yesterday, we checked the beam spot on the PR2 HR target, which is almost center (fig1): good.

After confirming the alignment is good, we checked beam at high power beam dump (fig2).
Beam path seems far enough from the beam dump: also good.

2. Check beam spot on in-vac POM behind PR2:

After confirming the IR alignment to PR2, we checked beam spot on in-vac POM behind PR2 (fig3)
It is hard to say from the picture, but no clip happens at in-vac POM in my eyes.

3. Reconstruction of POP forward beam path:

After confirming beam spot on in-vac POM is not so bad, we started reconstruction of POP forward beam path.
At upper mirror on the periscope, the beam is not center but it seems no clip (fig4), so we keeep it as it is.
Then, we moved upper periscope actuators and hit the beam on the mirror center of lower mirror on the periscope (fig3).
After that, we moved FST1 in JGW-T1909623-v11, because it is hard to avoid clipping without moving the mirror.
Then, we aligned the beam to lower right of the FST1 (fig6) to avoid clipping at the knob of the mirror mount.
Figure 7 shows the current beam spot near the knob, which seems not so bad in my eyes.
After that, we realigned all the downstream and centered the QPDs.

Note:

During the work, on of the actuators of lower mirror mount on the periscope hits the periscope itself and cannot rotate to CW direction (we can rotate it in CCW direction).
So, we cannot move up the beam spot on FST1 without moving periscope mirror upward.
Since I am not familiar with the periscope installed for the POP forward, I didn't lift it up and keep it as it is.

Images attached to this report
VAC (IX)
takashi.uchiyama - 15:09 Wednesday 24 April 2024 (29307) Print this report
Pumping down IXC
2024/04/24

Kimura, Uchiyama

We started pumping down IXC at 14:21 by the dry pump at IXV.
VAC (IX)
nobuhiro.kimura - 15:02 Wednesday 24 April 2024 (29306) Print this report
Comment to Pressurizing IXC to atmospheric pressure (29292)

[Kimura, M. Takahashi and Sawada (Hokuto)]
We restarted presuurization of IXC up to 9.8 x 10^4 Pa (~ atmospheric pressure) with G-2 class grade air on morning of 24th/Apr.
Detailes are as followes;
1. Re-start injection the G-2 grade air into IXC
 9:45 Start injection at 7.8 x 10^4 Pa
 10:23 8.3 x 10^4 Pa
 11:09 8.9 x 10^4 Pa
 11:45 9.5 x 10^4 Pa
 12:58 9.8 x 10^4 Pa.
 Stopped injection. 
2. After completion of pressurization to atmospheric pressure, repair of flange leak was done. k-log 29305
3. The amount of gas used for pressurization to 9.8 x 10^4 Pa from 2.3 x 10^2 Pas was 70 m^3.
The breakdown details of the gases used for the vacuum breaking of IXC are as follows:
 63 m^3 of high purity air (7 m^3 x 9 bottles, total 63 m^3)  and 7 m^3 of high purity nitrogen (7 m^3, 1 cylinder)  for a total of 70 m^3.

VAC (IX)
nobuhiro.kimura - 14:51 Wednesday 24 April 2024 (29305) Print this report
Comment to IXC vacuum leak test results (29267)

[Kimura and Sawada (Hokuto)]

 The leak at the IXC cross pipe flange (+Y side) was caused by forgetting to remove the tape that temporarily secures the elastomer seal.

 (See attached photos 1~6)
 The forgotten tape was compressed by the flange, and it is presumed that the leak occurred through a small gap created here.
The tape was removed and a new elastomer seal was attached to the gasket sealing surface to close the flange.
The claw clamps closing the flange (36 in total) was tightened diagonally with torques of 20, 40, and 60 Nm.

Images attached to this comment
MIF (General)
tomotada.akutsu - 7:22 Wednesday 24 April 2024 (29304) Print this report
Comment to Finalization of IFI-IMM-PRM: Day 3 PM; rough check around IFI and IMM (29283)

A modified drawing of 29285. We mis-understood that the nominal distance from the center of "the optical beam from PRM to PR2" and the center of the "PR2 HR mid baffle aperture" is about 6 mm according to JGW-T1910200-v4, while the correct nominal number of this center-to-center is 4.8 mm according to JGW-T2113078-v2. Hirata-san has confirmed 4.8 mm is correct center-to-center with CAD.

Images attached to this comment
LAS (High Power Laser R&D)
tomoyuki.uehara - 6:44 Wednesday 24 April 2024 (29303) Print this report
Comment to neoLASE high power laser I-P characteristic measurement (29300)
memo: Attached is I-P characteristics with linear fitteing. Fitted line (I: 4A-9A) is y = 8.0*x-20.2. In the region where the optical power increases linearly with the two amplifiers current, every 1A increases 8W in the optical power.
Images attached to this comment
MIF (General)
takafumi.ushiba - 22:45 Tuesday 23 April 2024 (29299) Print this report
Finalization of IFI-IMM-PRM: Day 4; Re-alignment of IMMT1, IMMT2, and PRM

[Ikeda, Hirata, Takahashi, Ushiba]

Abstract:

We moved PR2 about 3.3 mm in + Y direction and IMMT2 beam spot about 6 mm in + Y direction.
Then, we centered IR beam on PR2 and check beam spot on PRM, which is also close to the center.
So, alignment to PR2 seems fine now.

Detail:

1. Move PR2 in +Y direction:

For reducing GRX beam light power to identify GRX beam spot center easily, we requested ALS_FIBX guardian to SAFE state (GRX power becomes about 0.8 mW).
First, we moved PR2 so that GRX beam spot on PR2 HR target became nominal.
Since PR2 target was set as projecting PR2 center along IR injection beam (PRM-PR2 line), GRX should be off centered by 3.3 mm (165.48 mm (distance between PR2 HR surface and target) * 0.02 rad (angle between input and reflected beams of PR2)).
Figure 1 shows the GRX beam at PR2 HR target before moving PR2.
Figure 2 and 3 show the GRX beam spot after moving PR2 by 3.5 mm in +Y direction with the traverser.
Now, beam spot seem shifted in -Y direction more than 3 but less than 3.5 mm according to the fig3, which is almost the nominal value.

2. Move IMMT2 beam spot by moving IMMT1:

After moving PR2, we moved IMMT1 so that beam spot on IMMT2 shifted in + Y direction.
First, we requested PROVIDING_STABLE_LIGHT state to IO guardian and wait until alignment becomes good, and HOLD_ALIGNMENT state to IMC guardian not to kick the suspension when blocking the beam to IMMT1T QPDs.
Then, we set a ruler in front of IMMT2 and check the beam spot on the ruler before and after IMMT1 moves.
Figure 4 and 5 show the beam spot before and after moving IMMT1.
IMMT2 beam spot roughly moved 6 mm in +Y direction.

According to the current situation reported in klog29285, we need to shift PR2 beam spot by 5 mm along +Y direction.
Roughly speaking, 6 mm beam spot shift on IMMT2 moves PRM beam spot by 4.5 mm (6*3/4).
Due to PR2 position changes (3.3 mm along +Y direction), PRM beam spot will change 0.8 mm (3.3/4) in +Y direction when the beam is centered to PR2. 
So, in total, PRM beam spot changes by 5.3 mm, which is the same as we would like to move.

3. Center the beam to PR2 by moving IMMT2:

After moving IMMT1, we centered the IR beam on PR2 by moving IMMT2.
Then, we aligned PRM so that PRM reflection goes to REFL table.
Figure 6 shows the beam spot on PR2 target with aligned PRM, which seems well centered horizontally.
After that, we checked the beam spot on PRM AR target (fig7).
The beam seems hitting almost center.

Note:

This time, we haven't changed beam spot vertically though it is obviously higher than the nominal position on PR2.

Images attached to this report
Comments to this report:
naoatsu.hirata - 20:56 Wednesday 24 April 2024 (29312) Print this report

[Hirata, Ushiba-san, Ikeda-san]

We checked IR beam position around PR2 HR side Mid-size baffle. (How to make the vertical line is same as klog:29282)
It looks that IR beam positon is about 3.5mm +Y side away from the center of aperture.(IR beam center is 68.5mm and aperture center is 72mm on the ruler.)

Images attached to this comment
LAS (High Power Laser R&D)
tomoyuki.uehara - 18:49 Tuesday 23 April 2024 (29300) Print this report
neoLASE high power laser I-P characteristic measurement
[Miyakawa, Uehara]
After hosing the high power laser and power meter to the water chiller, the laser I-P characteristics of the laser were roughly measured. Attached is the IP characteristics. The laser power increased as the current of the amplifier increased. (The temperature setting of the water chiller is 20 degrees Celsius.)
Images attached to this report
Non-image files attached to this report
Comments to this report:
tomoyuki.uehara - 6:44 Wednesday 24 April 2024 (29303) Print this report
memo: Attached is I-P characteristics with linear fitteing. Fitted line (I: 4A-9A) is y = 8.0*x-20.2. In the region where the optical power increases linearly with the two amplifiers current, every 1A increases 8W in the optical power.
Images attached to this comment
DGS (General)
takahiro.yamamoto - 18:39 Tuesday 23 April 2024 (29302) Print this report
MEDM time machine is available

MEDM time machine works well.
It can show past values recorded in frame files on MEDM screens and can be launched from the MEDM execute menu.

After choosing a lock back time as relative time, gps time, local time etc. (see also the left windown in Fig.1), a new MEDM screen is launched. Channels not recorded in frame files such as _OUTMON are displayed as white boxes instead of record values as shown in the bottom window in Fig.1

Images attached to this report
VIS (IMM)
takafumi.ushiba - 18:02 Tuesday 23 April 2024 (29301) Print this report
IMMT1/2 OpLev centering and offload

After the alignment work (klog29299), we centered OpLevs and offload actuator outputs by using picomotor.
Followings are the steps we moved for offloading today.

IMMT1 pitch: -330 steps (400 -> 70)
IMMT1 yaw: +14950 steps (-2300 -> 12650)
IMMT2 pitch: +30 steps (-600 -> -570)
IMMT2 yaw: +9325 steps (-8500 -> 825)

VAC (EYA)
takashi.uchiyama - 16:27 Tuesday 23 April 2024 (29298) Print this report
Comment to Pool of liquid in EYA chamber (29214)
2024/04/23

Kimura, Uchiyama

We performed the pumping test again.
This time, we used a stronger dry pomp.

[Trial 1] No sample
The achieved vacuum pressure was 3*10^3Pa.

[Trial 2] with a liquid sample obtained last Friday. https://klog.icrr.u-tokyo.ac.jp/osl/?r=29280
We observed bubbles were generated but finally, the bubble generation stopped.
The achieved vacuum pressure was 3*10^3Pa.

We think that the origin of the bubbles was water.

[Trial 3] with the liquid sample used in the trial 2.
No bubble generation was observed during pumping down.
The achieved vacuum pressure was 3*10^3Pa.

The next time, we will use TMP.



Images attached to this comment
VAC (IX)
nobuhiro.kimura - 15:19 Tuesday 23 April 2024 (29296) Print this report
Comment to Pressurizing IXC to atmospheric pressure (29292)

[Kimura, M. Takahashi and Sawada (Hokuto)]
 We restarted presuurization of IXC up to 7.8 x 10^4 Pa with G-2 class grade air and G-2 class grade N2 on 23th/Apr.
Detailes are as followes;
1. Re-start injection the G-2 grade air into IXC
 8:34 Start injection at 3.2 x 10^4 Pa
 10:25 4.2 x 10^4 Pa
 11:34 5.5 x 10^4 Pa
 11:59 5.7 x 10^4 Pa
 13:06 6.6 x 10^4 Pa
 13:54 7.1 x 10^4 Pa 
 14:41 7.8 x 10^4 Pa.
 Stopped injection. 
2. Injection is resumed on the morning of April 24.
After completion of pressurization to atmospheric pressure, repair of flange leaks will begin.
The amount of gas used for pressurization to 7.8 x 10^4 Pa was pure air (7 m^3 x 7 bottles, total 49 m^3)
and pure nitrogen (7 m^3, 1 cylinder).

PEM (General)
shoichi.oshino - 15:14 Tuesday 23 April 2024 (29297) Print this report
Starting weather station PC
The weather station PC at Atotsu entrance was stopped on 4/12 due to the power supply inspection.
This PC was started on 4/23.
Search Help
×

Warning

×