with Dan, YamaT, SawadaT, Sembo
As noted in klog#33920, the HPCD described in klog#33854 was successfully installed. However, the transfer function measurements was unsuccessful. Investigation into the root cause was conducted, but no definitive reason could be identified.
After confirming that no abnormal voltage signals were present in the circuit or measurement setup, we replaced the SR560 which had been showing unusual behavior. A measurement equivalent to that shown in Fig.1 was then performed (Note that the coherence shown is in the range 0.999 – 1.0). As illustrated in Fig.2, the output signal (red trace in the lower panel) was clearly improved. This suggests that the issue was likely not due to the circuit itself.
Note:
Measurement equipment such as the SR560, spectrum analyzer, and related instruments used in this measurements were brought back to the analysis building.
Please refer to klog#33823 for the measurement setup and another example of HPCD transfer function measurements.