MIF (General)
hirotaka.yuzurihara - 20:06 Monday 28 April 2025 (33585) Print this report
lockloss investigation: 2025/04/21~04/27

I performed the lockloss investigation for the recent lockloss from the OBSERVATION state of the LSC_LOCK guardian, between 2025-04-22 10:23:43.062500 UTC and 2025-04-27 23:58:24.687500 UTC. The previous lockloss investigation was posted in klog33509. During this period, there were 22 lockloss.

Lockloss with the interferometer work

15 lockloss occurred in the WORKING_WITH_OBSERVATION_WITHOUT_LINES state.

1~10Hz seismic motion made the IMC lockloss

For 3 lockloss, we saw the excess of the seismic motion with 1~10 Hz, which made the oscillation of the IMC length control and made the IMC lockloss.
The time and date are summarized below:

Lockloss with unknown cause

For 4 lockloss, I'm not sure the lockloss cause.

They have common features just before the lockloss.

  • K1:ASC-DHARD_P_OUT_DQ reached the large negative value (such as  -400 ~ -600).
  • OMMT2 TRANS increased in the last a few seconds.

In the case of 2025-04-23 08:53:46.312500 UTC, the ASC DSOFT P and Y looks the oscillation with 0.07 Hz (Fig1).

Memo

Although this is not related to the lockloss cause, I found the transient change in IMMT1 oplev P and Y. All four segment detected same transient, as shown in Figure 2 and 3. Is that the glitch due to the oplev light source?