Reports 1-1 of 1 Clear search Modify search
VIS (IFI)
dan.chen - 17:35 Tuesday 25 February 2025 (32796) Print this report
Measurement of the Resonance of the IFI Steering Mirror Mount

With Yamazaki-san,

We are investigating whether the resonance of the IFI steering mirror could be the source of jitter noise around 120 Hz. In the previous measurement by Mitsuhashi et al. (link), a different type of mirror mount was used instead of the one currently in operation. This time, we plan to perform measurements using the same type of mount as in actual use.

Preparation Work (Today’s Tasks)

To conduct the measurement, we carried out the necessary preparations today:

  1. Laser Displacement Meter (LV-9300A):

    • We confirmed the proper usage of the laser displacement meter at the Mozumi office.
  2. Equipment Transport:

    • All required components were transported into the underground site.
  3. Fixture Preparation & Assembly:

    • A fixture was prepared to secure the laser displacement meter at the appropriate height.
    • The fixture was cleaned by wiping and then assembled.
  4. Measurement Location Setup:

    • Due to concurrent work around the OMC, the measurement will be conducted in the available space on the POS optical table.
    • Upon checking the table, we found many screws and optical components scattered around, so we tidied up:
      • Items that could be safely stored in a box were placed in a box and stored under the POS table.
      • Delicate optical components were gathered at the edge of the POS table.
  5. Fixture and Instrument Installation:

    • The fixture for securing the laser displacement meter was installed on the optical table.
    • The laser displacement meter was then mounted onto the fixture.
  6. HEPA Filter Operation:

    • During work on the POS optical table, while the side plate was open, two HEPA filters on the upper table were switched ON.
    • After completing the work, the side plate was closed, and the HEPA filters were switched OFF.

Next Steps

Tomorrow, we will begin the actual measurement.

Images attached to this report
Comments to this report:
dan.chen - 16:45 Wednesday 26 February 2025 (32805) Print this report

With Ikeda-san

We set up the same mirror mount as the STM inside the IFI chamber and started the measurement, but no significant peak was observed around 120 Hz. Further details will be reported later.

dan.chen - 8:38 Thursday 27 February 2025 (32820) Print this report
dan.chen - 10:51 Thursday 27 February 2025 (32819) Print this report

With Ikeda-san

We examined the new mirror mount (8822-AC-UHV). It appears that there are two possible ways to secure a mirror in this mount:

  1. Using the spring-loaded pawls at the back of the mirror.
  2. Using a 5/64 screw on the side of the mount.

The spring-loaded pawls are a unique mechanism that is not commonly seen in other mounts. These components, possibly made of PEEK, press the mirror from the back. Each pawl has a small bolt, but the pawl can be lifted and rotated without turning this bolt, allowing for easy retraction when inserting the mirror.
According to the manufacturer’s webpage:
"Spring-loaded pawls gently, but securely, hold a high-precision 2-inch (50.8 mm) diameter optic from the back against three points on the front plate of the mount."
This suggests that securing the mirror solely with the pawls is also a viable option. A conceptual diagram of the pawl mechanism is attached, illustrating my interpretation of its structure.

The actual method used to fix IFI STM mirrors inside the IFI chamber is unknown. To investigate, we performed vibration measurements under different fixation methods. The mirrors used in this test were the same as those from the previous measurement.
According to klog 18501, the actual IFI STM mirror is a 2-inch mirror with a thickness of 11 mm. The thickness of the mirror used in this test has not yet been verified.

Measurement Setup

The mirror mount was attached to the POS table using a pedestal post secured in three directions. We conducted vibration measurements under four different fixation methods:

  1. Side screw + three rear pawls
  2. Side screw only
  3. Side screw + three rear pawls (with one pawl screw pushed in by ~1.3 mm to mimic the actual STM2 fixation method)
  4. Three rear pawls only (with one pawl screw pushed in by ~1.3 mm to mimic the actual STM2 fixation method)

For each fixation method, we took two types of measurements:

  • "Free" measurement (no external excitation)
  • "Shake" measurement (lightly tapping the pedestal post from the back)

We plan to conduct additional measurements with different fixation methods in the future.

Results

No significant peak near 120 Hz was observed, and no major differences were found between the different fixation methods. However, minor differences around 600 Hz may exist.
The attached graph overlays all measurement results, using the side screw + rear pawl fixation method as the reference, which is assumed to be the standard fixation method..

Pictures

Images attached to this comment
dan.chen - 17:10 Thursday 27 February 2025 (32825) Print this report

With Hido-san,

We investigated how resonance changes depending on different methods of fixing mirror.

We found that the mirrors we had been using were 12 mm thick, so we installed and measured 11 mm thick mirror as well.

As a result, no resonance was observed around 120 Hz, and probably none around 220 Hz either (details to be provided later). A significant resonance was observed at 270 Hz. Upon investigation, we suspect that this resonance may be due to the "neck" of the pedestal.

A detailed report will follow.

dan.chen - 13:38 Friday 28 February 2025 (32847) Print this report

With Hido-san

Objective

The objective of this experiment was to complete the remaining measurements from the previous day and evaluate a newly identified mirror that matches the one inside the IFI chamber. Specifically, the focus was on determining whether different fixation methods affect the resonance frequency, checking for peaks around 120Hz and 220Hz, and investigating the cause of the prominent peak observed at 270Hz.

Procedure

1. Measurements with Variations

  • Conducted measurements under different conditions:
    • With and without shaking.
    • With shaking of the laser source.
    • With the side screw released or tightened.
    • With the pawls in different configurations:
      • Released.
      • Restored to the original position.
      • Fully tightened.
      • Adjusted to simulate the IFI chamber conditions.

Results and Considerations

  • No significant changes were observed regardless of the fixation method, indicating that the 120Hz and 220Hz noise sources are likely unrelated to the mirror mounting configurations.

2. Investigation of 270Hz Resonance

  • Applied force using a wrench to assess its effect on resonance behavior.
  • Inserted a washer between the mirror holder and the mount to study its influence on resonance characteristics.
  • Attached the mirror mount on a thin post and conducted measurements.

Results and Considerations

  • The 270Hz peak remained unchanged when force was applied to the thick pedestal, and it disappeared when the mirror was mounted on a thin pole, suggesting that the source of resonance is neither the pedestal itself nor the mirror mount.
  • Introducing a washer between the mirror mount and the pedestal caused some variation in the peak, leading to the hypothesis that the neck of the pedestal might be the primary source of the 270Hz resonance.

3. Testing Different Mirror Mount Configurations

  • Initially, measurements were conducted using a 12mm mirror.
  • Later, an 11mm mirror(link), identical to the one inside the IFI chamber, was found and tested under the same conditions.
  • The PYD-20 mirror was examined but not measured due to its significantly different thickness.
  • Adjusted the mirror’s positioning to simulate IFI chamber conditions.

4. Thickness Measurements

  • Measured the thickness of the 12mm and 11mm mirrors.
  • The PYD-20 mirror was inspected but not measured.

Conclusion

The various configurations and shaking tests were performed to assess the mechanical stability and potential contributions to noise. The analysis of the collected data showed no peaks around 120Hz and 220Hz. The prominent peak observed at 270Hz is likely due to the resonance of the pedestal’s neck region.

Legend for Attached Graphs

  • "Back: locked(+1.9mm)": Indicates that the back pawls' screws were adjusted to protrude 1.9mm from the front.
  • "Back: locked(only one pawl)": Denotes a condition where only one of the three back pawls was locked while the other two were released, meaning the mirror was held in a very loose state.
Images attached to this comment
dan.chen - 13:40 Friday 28 February 2025 (32848) Print this report

As there are too many graphs, I picked up the data in the case of 11mm thickness mirror and related graphs.

Images attached to this comment
dan.chen - 13:50 Friday 28 February 2025 (32849) Print this report
dan.chen - 6:10 Saturday 01 March 2025 (32857) Print this report

Slides used in a commissioning meeting for this topic: JGW-T2516543

dan.chen - 17:44 Monday 03 March 2025 (32879) Print this report

With Kenta Tanaka

We first attempted to reproduce the same conditions as last week and conducted measurements. As a result, a peak appeared at 120 Hz. The cause of this peak is unknown, but it persisted even after changing the mirror mounting method. Electrical noise is also a possible factor.

To investigate further, we inserted a small optical breadboard plate (MB1515/M) under the pedestal. We then placed the pedestal at the edge of the plate and performed measurements. A broad peak noise around 120 Hz was observed, which varied depending on the mounting method (using two clamps) and the amount of overhang.

In the final measurement, we kept the clamping method as consistent as possible and only varied the overhang. As a result, the peak frequency changed.

Graphs:

  • First graph: The 120 Hz peak observed under the same conditions as last week, without the plate.
  • Second graph: The difference in the final measurement results. A comparison between cases where the overhang was 4.6 mm and 6.3 mm.

Detailed data will be reported later.
The POS table has been neatly cleaned up. All measurement equipment, mounts, and so on have been brought back to the office, making it possible to conduct additional tests at Toyama University or other locations if needed.

Images attached to this comment
Search Help
×

Warning

×