I updated the reference plots to the recent healthy ones (2024/03/22, after oplev centering, klog28942 ).
The PIT resonance frequecy around 3.5 Hz seems to shift lower, and DC gain slightly increased. The others seem to fine (H2 and H4 measurements seem to be performed at a noisy timing?)
With Ohshino-san, Ushiba-san
The latast measurements look healthly.
The difference from the reference data and the reason we think.
At first, we checked the actuator gain (COILOUTF gain) for each coil. The values changed from the reference measurement time.
We think this caused the difference we found.
With Oshino-san
We performed the off-load work for MCE.
Procedure
After this work, we requested LOCK_AQUISITION to check. It works.
Then, we kicked health check script.