takahiro.yamamoto - 15:06 Sunday 24 November 2024 (31747) Print this report
Comment to Re-tuning of PLL temperature bias (31744)

> Is it similar behaviour reported in klog30314?
Yes to the fact that PLL lock by guardian is prevented by the jump of IMC temperature feedback. 

No to the reason why IMC temperature feedback jumped.
In the previous case, IMC guardian executed "clear history" because the feedback output became lager than software threshold. As a result, IMC temperature feedback jumped to 0ct. In this time temperature feedback jumped from ~-5000ct to ~-6000ct. So it's not due to "clear history".

In any case, PLL guardian must become to find a proper temperature bias on its own.