Reports 1-1 of 1 Clear search Modify search
VIS (OMM)
ryutaro.takahashi - 8:12 Saturday 16 November 2024 (31659) Print this report
Comment to OMMT1 recovery work (31633)

(Log on 14th)

I investigated the glitches (Fig.1) in the H2 OSEM of the OMMT1 suspension. When the D-sub cable was disconnected from the feedthrough, the glitches disappeared. The LED's voltage drop in forward bias decreased.

  7/10 11/14
1-6: PD 0.704V 0.710V
2-7: LED 1.191V 0.954V
3-8: Coil 16.8Ω 16.2Ω


I found the shield lines in the D-sub cables for H2 and H3 disconnected from the satellite box (arrows in Fig.2). Connecting the shield lines reduced the high-frequency noises (Fig.3), but the glitches did not disappear.

Images attached to this comment
Search Help
×