Here is the link to the LIGO DCC for the circuit design of the QPD transimpedance amplifier.
https://dcc.ligo.org/LIGO-D1001974
KAGRA uses basically the same design as above.
The noise subtraction path can be enabled/disabled by moving the jumper pin (P1).
It was disabled before.
We can adjust the subtraction gain by changing R13 of the noise cancellation amplifier.
Since the stationary noise level is better with the noise subtraction turned off (even with the optimized values of R13), we left it turned off for the moment.
We have two issues with this circuit.
1. The stationary noise is worse with the noise subtraction enabled.
2. We see glitches. They are sporadic and elusive.
The glitch issue is more serious than the small stationary noise increase.
We have only tested long operation with the noise subtraction disabled so far. So we will enable the noise subtraction and leave it for a while to see if the glitch problem is eased by the noise subtraction or not.
Yokozawa, Aso
We turned on the noise subtraction path in the OMC QPD driver.
We also tried to isolate the chasis from the rack by adding insulation between the bracket and the rack as shown in the attachment no.3.
We confirmed that the chasis is isolated from the rack in this state with a tester.
However, when we connected cables, the chasis was electrically connected to the rack. Especially, the cables between the QPD driver and the AA chasis connect the two chasis through the shell of the D-Subs.
After resinstallation of the driver into the rack, the output of the driver moved around a lot (a few tens of counts). This is a well known behavior of this circuit, probably due to the temperature drift.
We waited for ~30min, and the strange behavior of the output signals went away.
The attachment no.4 is the QPD noise spectra after the above wait.
From 12:30 to 13:30 (JST) of yesterday, the laser was shutdown, thus OMC QPDs did not get any light.
I checked the time series of the QPD outputs during this period.
Obviously, the outputs move around in the order of 0.2 counts.
Note that these signals are normalized by the sums of the QPDs. Usually, they get more than 10000 counts in sum, but it was order of 10 when no light falls on the QPDs.
Therefore, 0.2counts of noise becomes order of 1e-4 or less counts in operation. Probably this is not a big issue for initial alignment.
Zooming into a glitch, they are pretty fast change of the offset. We can see some transient response from the circuit (probably the whitening filter).
The cause of these glitches are not understood at this moment.