[KTanaka, YokozaWashimi]
Today we aligned the new D-sub9 cables for
- OMC DCPD
- OMC QPD1
- OMC QPD2
- OMC PZT
- Z switch
- spare1
- spare2
using the new cable lack.
After this work, we checked them. But we found
- OMC DCPD A,B : Still noisy
- OMC QPD1,2: The driver oscillation problem (?) happened again and we couldn't check the cabling noise.
- OMC PZT: We couldn't check it without laser.
- Z switch: worked
(Deatails will be reported later)
The today's final configuation of these cables are:
- OMC DCPD : connected to the floor cable
- OMC QPD1,2: disconnected
- OMC PZT: connected by the new cable
- Z switch: connected by the new cable