satoru.takano - 15:48 Friday 26 July 2024 (30588) Print this report
Comment to Consideration of OMC trans noise caused by OMC length noise transferred from the OMC chamber optics table (30498)

Supplemental material

Response from the displacement (x) of the cavity to the relative intensity (RIN):

RIN(x) = \frac{1}{1 + \left(\frac{x}{HWHM}\right)^2} \simeq 1 -\left(\frac{x}{HWHM}\right)^2

Fluctuation in a certain frequency:

RIN(A\cos\omega t) \simeq 1 -\left(\frac{A\cos\omega t}{HWHM}\right)^2 = 1 - \frac{1}{2}\left(\frac{A}{HWHM} \right)^2 - \frac{1}{2}\left(\frac{A}{HWHM} \right)^2 \cos2\omega t

Amplitude of response at the second harmonics of the original frequency:

RIN \,at \,2\omega = - \frac{1}{2}\left(\frac{A}{HWHM} \right)^2 = -2\left(\frac{A}{FWHM} \right)^2