MIF (General)
satoru.takano - 7:48 Wednesday 24 July 2024 (30539) Print this report
Comment to OMC calibration by gradient of the error signal (30517)

I took scanning data with the following settings:

  • The input of OMC-LSC_NORM is turned off
  • The offset of OMC-LSC_NORM is set to 30 (in the calculation finally this value was corrected by the actual value at the peak)
  • Dither amplitude is set to 300
  • ISS is engaged

The scanning result is shown in the Figure1. With this timeseries data, I calculated the gradient of the error signal and got the calibration factor of 6.4(5)e9 [cnt/m]. This value is a bit larger than the value obtained by the transfer function measurement (5.1(4) e9 [cnt/m]), but seems consistent.