hirotaka.yuzurihara - 15:54 Tuesday 09 July 2024 (30321) Print this report
Comment to IRX spectrum characterization (30110)

I performed the crosscheck of the input quivelent noise of the input for the CARM CMS, when the IRX is locked (with Ushiba-san's help). I didn't find the critical mistake.

  • I checked the pole and zero of the generic filters (wiki page) used in CARM CMS.
  • I checked if the filter bank at the measurement time worked for the dewhitening filter and if the necessary FM switch was on.
  • I derived the gain factors from the DAQ values to input of the CARM CMS and confirmed they are correct.
  • I checked if the pole and zero for the calibtation from the DAQ to input of the CARM CMS is correct or not.