Date: 2024/7/1
I performed WSK/GSK calibration at ToyamaU with the new automation system.
The main result is the following.
The difference from the results of O4a is about 2%, which was 0.142.
- The reason is not clear.
- Investigation I did today:
- Cable effect: not observed for GSK and WSK.
- Beam position at port: no effect.
- Incident angle could not be changed.
- Both yaw and pitch angle should be better than O4a.
Detailed procedure
- Move WSK and GSK to be "formation A".
- Measure the dark noise (N=2000)
- Measure the main data (N=200)
- Move WSK and GSK to be "formation B".
- Measure the dark noise (N=2000)
- Measure the main data (N=200)
- Convert the result to calc alpha_WSKpGSK.
- Repeat 1-3 for 10 times.
Detailed report: link