Reports 1-1 of 1 Clear search Modify search
VIS (BS)
ryutaro.takahashi - 22:40 Wednesday 08 May 2024 (29432) Print this report
Comment to Health check of BS (28292)

I checked the TF measured in air. The results were consistent with the last measurement in vacuum. The Q of the GAS filters seems smaller than the last measurement (The script for the BF GAS was not working). The TF of the IM H3 OSEM, in which the flap was rotated more than 40°, is consistent with the reference.

Images attached to this comment