CRY (Cryo-payload EX)
takafumi.ushiba - 17:36 Friday 19 January 2024 (28334) Print this report
Apply current to the IM heater on EX payload

Since we started the payload cryocooler to keep the radiation shield temperature low (klog28275), EX payload temperature is now gradually decreasing.
So, I turned on IM heater on EX payload to keep the current temperature.

Setup of power supply can be seen in fig 1.

Images attached to this report
Comments to this report:
shinji.miyoki - 8:47 Saturday 20 January 2024 (28342) Print this report

The MN temp went from 78K to 88K at 8:40.

shinji.miyoki - 9:20 Monday 22 January 2024 (28347) Print this report

The temp is now 102K and on the process of increasing. Ushiba-kun will reduce the current.

takafumi.ushiba - 9:47 Monday 22 January 2024 (28349) Print this report

I reduced the current from 0.1 to 0.05 A (fig1).

Images attached to this comment
shinji.miyoki - 9:23 Tuesday 23 January 2024 (28367) Print this report

94K around 9:30, and on the way of reduction.

Images attached to this comment
takafumi.ushiba - 16:34 Wednesday 24 January 2024 (28398) Print this report

I increased IM heater current to 0.2 A (fig1).

Images attached to this comment
takafumi.ushiba - 17:44 Friday 09 February 2024 (28566) Print this report

I turned off ETMX IM heater.

satoru.ikeda - 12:57 Friday 07 March 2025 (32919) Print this report

Ushiba-san(remote), Ikeda

Around 10:00, We turned on IM heater of ETMX(fig1).

Images attached to this comment
shinji.miyoki - 15:40 Friday 07 March 2025 (32926) Print this report

The IM temp is increasing.

Images attached to this comment
takafumi.ushiba - 10:28 Tuesday 11 March 2025 (32951) Print this report

I reduced the current applied to IM heater from 0.1 A to 0.08 A (fig1).

Images attached to this comment
shinji.miyoki - 10:46 Tuesday 11 March 2025 (32954) Print this report

The temps of EX and EY IM are almost the same, ~65.6K around 10:44. However, the currents for IMs are almost the same: 0.08A and 0.078A. So the current for EY of 0.08A is better to reduce because of the lower cooling power for the EY payload.

shinji.miyoki - 3:47 Wednesday 12 March 2025 (32966) Print this report

The present IM temperatures of EX and EY. So, should the current for the EX IM heater be more than 0.08A, but less than 0.1A? On the other hand, the current for the EY IM heater should be kept?

Images attached to this comment
takafumi.ushiba - 10:33 Tuesday 25 March 2025 (33091) Print this report

I increased the current for EX IM heater from 0.08A to 0.088A (fig1).

Images attached to this comment
shinji.miyoki - 7:57 Wednesday 26 March 2025 (33104) Print this report

At least 90K can be acceptable for the IM temp. However, the warmup speed seems to be small.

Judging from the temp situations just before starting cryocoolers for the payload, only outer shield temps (80K_SHIELD_SIDEA/B/TOP) and the 4K_REF2/4_50K_HEAD, and maybe the area between these two points are over 157K at which the water will come out.

In addition, the vacuum level changes at EXT and EXC were small (8->5x10^-6Pa, 5->2.5x10^-6Pa) when the temperatures of these areas fell below 157K. this means that the water reduction and increase effect will be small.

So, we believe we can stop these cryocoolers for payloads (or one of the coolers) for a while to accelerate the warmup. However, cracking noise could be occur again for a while.

Images attached to this comment
takafumi.ushiba - 9:26 Wednesday 26 March 2025 (33106) Print this report

I increased the current for EX IM heater from 0.088A to 0.102A (fig1).

Images attached to this comment
shinji.miyoki - 7:30 Thursday 27 March 2025 (33114) Print this report

I increased the current for EX IM heater from 0.110A to 0.102A (fig1).

Images attached to this comment
shinji.miyoki - 7:44 Sunday 30 March 2025 (33160) Print this report

I reduced the current for the EX IM heater as Fig.1.

Images attached to this comment
shinji.miyoki - 6:57 Tuesday 01 April 2025 (33181) Print this report

I reduced the current from 0.109 A to 0.107 A.

Images attached to this comment
shinji.miyoki - 9:56 Thursday 03 April 2025 (33204) Print this report

I increased the current from 0.107 A to 0.108 A.

Images attached to this comment