CRY (Cryostat EX)
nobuhiro.kimura - 20:16 Thursday 18 January 2024 (28324) Print this report
Comment to EX ref3 cryocooler was stopped around 18:40 on January 14th. (28275)

[Kimura, M.Takahashi and Sawada]
 We replaced the failed Ref 3 (P-55) compressor with a spare and restarted cooling of the inner radiation shield.
We conducted a simple inspection by irradiating a black light on the discharge side connection of the removed compressor and the connection piping.
 As a result, luminescence indicating oil traces were observed at the gas contact area. Therefore, the connection piping and filter from Ref 2 (P-54) compressor were converted to Ref 3 (P-55).
The connecting piping and filter need to be newly replaced.
Currently, there are a total of five cryo-coolers in operation at EXC: two duct shielded cryo-coolers, Ref1 (P-53) for radiation shield , Ref3 (P-55)for radiation shield, and Ref4 (P-56)  for payload.