CRY (Cryostat IX)toshiro.ochi - 2:44 Wednesday 01 March 2017 (2561)
Print this reportToday's result of measurement vibration of six cryocoolers(full-power mode!!)[Ochi,K.Yamamoto]
Today, we measured vibration of six cryocoolers using cryo-accerelometer installed on the cryostat inner shield.
This result(graph) shows the peaks of 30-40 Hz are very big(~1.0e-07 m/rtHz).
It amplitude is the same of Dan Chen's result (measured at Toshiba Keihin Office).
And result of Trillium-Speedmeter put near the Cryostat shows similar to cryoacc's.
(Attention, result of cryocoolers on of cryoacc is too noisy so I cannot measure the vibration on 1-10 Hz accurately.)
Images attached to this report
Comments to this report:
kazuhiro.yamamoto - 22:33 Saturday 04 March 2017 (2572)
Print this reportProgress in the last one week.
27th of Feb (Mon)
all cryocoolers (four cryocoolers for cryostat and two cryocoolers for cryoduct) started to work
(16:50 JST). We measured vibration before and after start of cooling. We observed cryocooler generates
vibration. Details are being analyzed.