PEM (Center)takaaki.yokozawa - 8:13 Thursday 11 May 2023 (25190)
Print this reportAcoustic injection to POP area 230511I performed the acoustic injection test to POP area this morning.
50 count injection 07:27:30 - 08:01:00 (JST) 50 - 900 Hz with 5 Hz resolution 10 second acoustic injection in one frequency
Fig.1. showed the whitened spectrogram for POP table microphone Fig.2. showed the whitened spectrogram of DARM displacement As you can see, when we injected the sound around 400 Hz(Fig.3) and 360 Hz(Fig.4.) large excess can be obtained.
Images attached to this report
Comments to this report:
tatsuki.washimi - 9:56 Thursday 11 May 2023 (25193)